Leica EL6000 – Specifications (cont'd.)

Light output

Design

Type: Storz, long

Special feature

Self-locking if no light guide is

 

plugged in

Fan

 

 

Switch-on delay

Approx. 60 s after lighting the

 

lamp

 

Interface

 

 

Type

Remote control input for

 

internal­

shutter

Design

9-pin D-Sub socket

Remote signal level

 

 

–  Shutter closed

–8 to +0.8 V

–  Shutter open

+2.4 to +8 V

–  Input current

Approx. 1 mA

Footswitch

 

 

–  Shutter closed

Connection between

 

pin 2 and 4

–  Current

Approx. 1.2 mA

Shutter frequency

Max. 10 Hz

Protective functions

Safety circuit

If the cover is open, all components in the instru-

 

ment (except for the power input and power plug)

 

are disconnected from the power supply voltage.

Overheating

In the event of overheating, the lamp supply is shut

protection

off. After cooling off, the lamp power supply in the

 

Leica EL6000 compact light source is automatically

 

switched on again.

Protection screen

If no light guide is plugged into the light output of

 

the instrument, the light output is automatically

 

closed off so that no light can escape.