CALibrate:PASSword

This command can only be used in calibration mode. It allows you to change the calibration password. A new password is automatically stored in nonvolatile memory and does not have to be stored with CALibrate:SAVE. If the password is set to 0, password protection is removed and the ability to enter the calibration mode is unrestricted.

Command Syntax

CALibrate:PASSword <NRf>

Parameters

0 (default)

 

Examples

CAL:PASS 6812

CAL:PASS 02.1997

Related Commands

CAL:STAT

 

CALibrate:PWM:FREQuency

Agilent 6811B, 6812B, 6813B Only

This command is only used during manufacture or repair. It trims the switching frequency of the power output stages. The numbers from 0 to 7 are internally mapped to 8 discrete frequencies.

Command Syntax

CALCulate:PWM:FREQuency <NRf>

Parameters

0 through 7

Examples

CAL:PWM:FREQ 1

Query Syntax

CALibrate:PWM:FREQuency?

Returned Parameters

<NR1>

Related Commands

CAL:PWM:RAMP

CALibrate:PWM:RAMP

Agilent 6811B, 6812B, 6813B, Only

This command modulates the slope of voltage ramp driving the power output stages. Varying the ramp affects the harmonic distortion of the output. The argument is a number from 0 to 255. This command is only used during manufacture or repair of the .

Command Syntax

CALCulate:PWM:RAMP <NRf>

Parameters

0 through 255

Examples

CAL:PWM:RAMP 100

Query Syntax

CALibrate:PWM:RAMP?

Returned Parameters

<NR1>

Related Commands

CAL:PWM:FREQ

CALibrate:SAVE

This command can only be used in calibration mode. It saves any new calibration constants (after a current or voltage calibration procedure has been completed) in nonvolatile memory.

Command Syntax

CALibrate:SAVE

Parameters

None

Examples

CAL:SAVE

Related Commands

CAL:CURR CAL:VOLT CAL:STAT