Agilent Technologies 6811B, 6834B, 6814B MEASure CURRent NEUTral AC?, FETCh CURRent NEUTral AC?

Models: 6811B 6813B 6834B 6843A 6814B

1 187
Download 187 pages 38.86 Kb
Page 46
Image 46

3 - Language Dictionary

MEASure:CURRent:NEUTral:AC?

FETCh:CURRent:NEUTral:AC?

Agilent 6834B Only

These queries return the ac rms current in the neutral output terminal of a three-phase ac source.

Query Syntax

MEASure:[SCALar]:CURRent:NEUTral:AC?

 

FETCh:[SCALar]:CURRent:NEUTral:AC?

Parameters

None

 

Examples

MEAS:CURR:NEUT:AC?

FETC:CURR:NEUT:AC?

Returned Parameters

<NR3>

 

Related Commands

MEAS:CURR:NEUT? MEAS:CURR:NEUT:ACDC?

MEASure:CURRent:NEUTral:ACDC?

FETCh:CURRent:NEUTral:ACDC?

Agilent 6834B Only

These queries return the ac+dc rms current in the neutral output terminal of a three-phase .

Query Syntax

MEASure:[SCALar]:CURRent:NEUTral:ACDC?

 

FETCh:[SCALar]:CURRent:NEUTral:ACDC?

Parameters

None

 

Examples

MEAS:CURR:NEUT:ACDC?

FETC:CURR:NEUT:ACDC?

Returned Parameters

<NR3>

 

Related Commands

MEAS:CURR:NEUT? MEAS:CURR:NEUT:AC?

MEASure:CURRent:NEUTral:HARMonic?

FETCh:CURRent:NEUTral:HARMonic?

Agilent 6834B Only

These queries return the rms amplitude of the Nth harmonic of current in the neutral output terminal of a three-phase ac source.

The parameter is the desired harmonic number. Queries sent with a value of 0 return the dc component. A value of 1 returns the fundamental output frequency. Harmonic orders can be queried up to the fundamental measurement bandwidth of the measurement system, which is 12.6kHz. Thus the maximum harmonic that can be measured is dependent on the output frequency. Any harmonics that represent frequencies greater than 12.6kHz are returned as 0.

Query Syntax

MEASure:[SCALar]:CURRent:NEUTral:HARMonic[:AMPLitude]? <NRf>

 

FETCh:[SCALar]:CURRent:NEUTral:HARMonic[:AMPLitude]? <NRf>

Parameters

0 to 50

 

Examples

MEAS:CURR:NEUT:HARM? 3

FETC:CURR:NEUT:HARM? 1

Returned Parameters

<NR3>

 

Related Commands

MEAS:CURR:NEUT:HARM:PHAS?

46

Page 46
Image 46
Agilent Technologies 6811B, 6834B MEASure CURRent NEUTral AC?, FETCh CURRent NEUTral AC?, MEASure:CURRent:NEUTral:ACDC?