Execution Errors –200 through –299 (sets Standard Event Status Register bit #4)

C - Error Messages

–170

–171

–178

–200

–221

–222

–223

–224

–225

–270

–272

–273

–276

–277

–310

–350

–400

–410

–420

–430

–440

0

1

2

3

4

5

6

7

10

11 - 31

40

41

42

43

Expression error

Invalid expression

Expression data not allowed

Execution Errors –200 through –299 (sets Standard Event Status Register bit #4)

Execution error [generic]

Settings conflict [check current device state]

Data out of range [e.g., too large for this device]

Too much data [out of memory; block, string, or expression too long] Illegal parameter value [device-specific]

Out of memory

Macro error

Macro execution error

Illegal macro label

Macro recursion error

Macro redefinition not allowed

System Errors –300 through –399 (sets Standard Event Status Register bit #3)

System error [generic]

Too many errors [errors beyond 9 lost due to queue overflow]

Query Errors –400 through –499 (sets Standard Event Status Register bit #2)

Query error [generic]

Query INTERRUPTED [query followed by DAB or GET before response complete] Query UNTERMINATED [addressed to talk, incomplete programming message received] Query DEADLOCKED [too many queries in command string]

Query UNTERMINATED [after indefinite response]

Selftest Errors 0 through 99 (sets Standard Event Status Register bit #3)

No error

Non-volatile RAM RD0 section checksum failed

Non-volatile RAM CONFIG section checksum failed

Non-volatile RAM CAL section checksum failed

Non-volatile RAM WAVEFORM section checksum failed

Non-volatile RAM STATE section checksum failed

Non-volatile RAM LIST section checksum failed

Non-volatile RAM RST section checksum failed

RAM selftest

DAC selftest error, expected <n>, read <reading>

Errors 11, 12, 13, 14, 15 apply to DAC12 1A and 1B

Errors 16, 17, 18 apply to DAC12 2A

Errors 19, 20, 21 apply to DAC12 2B

Errors 22, 23 apply to DAC12 4A

Errors 24, 25 apply to DAC12 4B

Errors 26, 27, 28 apply to DAC12 3A and 3B

Errors 29, 30, 31 apply to DAC12 5A and 5B

Voltage selftest error, output 1

Voltage selftest error, output 2

Voltage selftest error, output 3

Current selftest error, output 1