Agilent Technologies 6812B, 6834B, 6814B, 6843A, 6813B, 6811B manual PULSe:HOLD, Puls:Hold Dcyc

Models: 6811B 6813B 6834B 6843A 6814B

1 187
Download 187 pages 38.86 Kb
Page 83
Image 83
PULSe:HOLD

Language Dictionary - 3

PULSe:HOLD

This command specifies whether the pulse width or the duty cycle is to be held constant when the pulse period changes. The following tables describe how the duty cycle, period, and width are affected when one, two, or all three parameters are set in a single program message.

Command Syntax

[SOURce:]PULSe:HOLD <parameter>

Parameters

WIDTh DCYCle

*RST Value

WIDTh

Examples

PULS:HOLD DCYC

Query Syntax

[SOURce:]PULSe:HOLD?

Returned Parameters

<CRD>

Related Commands

PULS:COUN PULS:DCYC PULS:PER PULS:WIDT

 

 

PULSe:HOLD = WIDTh

 

Parameter Set

DCYCle

 

PERod

WIDTh

 

 

 

Set

 

 

 

 

 

 

Set

 

 

 

 

 

 

 

Set

Set

 

 

 

 

Set

 

 

 

Set

 

 

Set

Set

 

Set

 

Set

 

Set

Set

 

 

 

 

 

 

 

 

Parameter Set

DCYCle

PERod

WIDTh

 

 

Set

 

Set

 

 

Set

Set

 

 

 

Set

 

 

Set

 

Set

Set

Set

 

Set

Set

Set

 

 

 

Action

Sets WIDTh. If WIDTh < PERiod, recalculates DCYCle; otherwise recalculates the PERiod and DCYCle.

Sets PERiod. If WIDTh < PERiod, recalculates DCYCle; otherwise recalculates the PERiod and DCYCle.

Sets WIDTh. If WIDTh < PERiod, sets the PERiod and recalculates DCYCle; otherwise recalculates the PERiod and DCYCle.

Sets DCYCle and recalculates the PERiod.

Sets DCYCle and WIDth and recalculates the PERiod.

Sets DCYCle and PERiod and recalculates the WIDTh.

Sets WIDTh. If WIDTh < PERiod, sets the PERiod and recalculates DCYCle; otherwise recalculates the PERiod and DCYCle.

PULSe:HOLD = DCYCle

Action

Sets WIDTh and recalculates the PERiod.

Sets PERiod and recalculates the WIDTh.

Sets WIDTh. If WIDTh < PERiod, sets the PERiod and recalculates DCYCle; otherwise recalculates the PERiod and DCYCle.

Sets DCYCle and recalculates the PERiod.

Sets DCYCle and WIDth and recalculates the PERiod.

Sets DCYCle and PERiod and recalculates the WIDTh.

Sets WIDTh. If WIDTh < PERiod, sets the PERiod and recalculates DCYCle; otherwise recalculates the PERiod and DCYCle.

83

Page 83
Image 83
Agilent Technologies 6812B, 6834B, 6814B, 6843A, 6813B, 6811B manual PULSe:HOLD, Puls:Hold Dcyc