CURRent:PEAK:TRIGgered

3 - Language Dictionary

CURRent:PEAK:TRIGgered

Agilent 6811B, 6812B, 6813B, Only

This command sets the output limit of the absolute value of peak instantaneous current when a step or pulse transient is triggered.

Command Syntax

Parameters

Unit

*RST Value

Examples

Query Syntax

Returned Parameters

Related Commands

[SOURce:]CURRent:PEAK:TRIGgered <NRf+> refer to Specifications Table in User’s Guide

A(peak amperes)

13 (Agilent 6811B/6812B)

26 (Agilent 6813B) CURR:PEAK:TRIG 15

[SOURce:]CURRent:PEAK:TRIG?

<NR3>

CURR:PEAK CURR:PEAK:MODE

CURRent:PROTection:STATe

This command enables or disables the overcurrent (OC) protection function. If the overcurrent protection function is enabled and the exceeds the programmed level, then the output is disabled and the Questionable Condition status register OC bit is set (see Chapter 4 under “Programming the Status Registers”). An overcurrent condition can be cleared with OUTPut:PROTection:CLEar after the cause of the condition is removed.

NOTE: Use OUTP:PROT:DEL to prevent momentary current limit conditions caused by programmed output changes from tripping the overcurrent protection.

Command Syntax

[SOURce:]CURRent:PROTection:STATe<bool>

Parameters

0 1 OFF ON

 

*RST Value

OFF

 

Examples

CURR:PROT:STAT

0CURR:PROT:STAT OFF

Query Syntax

[SOURce:]CURRent:PROTection:STATe?

Returned Parameters

0 1

 

Related Commands

OUTP:PROT:CLE

OUTP:PROT:DEL

64