Language Dictionary - 3

OUTPut:IMPedance

Agilent 6811B, 6812B, 6813B, Only

This command enables or disables the output impedance programming capability of the ac source.

Command Syntax

OUTPut:IMPedance[:STATe] <bool>

Parameters

0 1 OFF ON

 

*RST Value

OFF

 

Examples

OUTP:IMP 1

OUTP:IMP ON

Query Syntax

OUTPut:IMPedance[:STATe]?

Returned Parameters

0 1

 

Related Commands

OUTP:IMP:REAL

OUTP:IMP:REAC

OUTPut:IMPedance:REAL

Agilent 6811B, 6812B, 6813B, Only

This command sets the real part of the output impedance of the ac source. OUTPut:IMPedance:STATe must be enabled for the programmed impedance to affect the output.

Command Syntax

OUTPut:IMPedance:REAL <NRf>

Parameters

0 to 1 (ohms)

*RST Value

0

Examples

OUTP:IMP:REAL 0.25

Query Syntax

OUTPut:IMPedance:REAL?

Returned Parameters

<NR3>

Related Commands

OUTP:IMP OUTP:IMP:REAC

OUTPut:IMPedance:REACtive

Agilent 6811B, 6812B, 6813B, Only

This command sets the reactive part of the output impedance of the ac source. OUTPut:IMPedance:STATe must be enabled for the programmed impedance to affect the output.

Command Syntax

OUTPut:IMPedance:REACtive <NRf>

Parameters

0.00002 to 0.001 (henrys)

*RST Value

0.0005

Examples

OUTP:IMP:REAC 100E-6

Query Syntax

OUTPut:IMPedance:REAC?

Returned Parameters

<NR3>

Related Commands

OUTP:IMP OUTP:IMP:REAL