Agilent Technologies 6812B manual Source Subsystem Frequency, FREQuency:MODE, Subsystem Syntax

Models: 6811B 6813B 6834B 6843A 6814B

1 187
Download 187 pages 38.86 Kb
Page 65
Image 65
Source Subsystem (Frequency)

Language Dictionary - 3

Source Subsystem (Frequency)

This subsystem programs the output frequency of the ac source.

Subsystem Syntax

[SOURce:]

 

FREQuency

 

[:CW :IMMediate] <n>

Sets the frequency

:MODE <mode>

Sets frequency mode (FIX STEP PULS LIST)

:SLEW

 

[:IMMediate] <n> INFinity

Sets the frequency slew rate

:MODE <mode>

Sets frequency slew mode (FIX STEP PULS LIST)

:TRIGgered <n> INFinity

Sets the triggered frequency slew rate

:TRIGgered <n>

Sets the triggered frequency

FREQuency

 

This command sets the frequency of the output waveform.

Command Syntax

[SOURce:]FREQuency[:CW :IMMediate] <NRf+>

Parameters

refer to Specifications Table in User’s Guide

Unit

HZ (hertz)

*RST Value

60

Examples

FREQ 50

Query Syntax

[SOURce:]FREQuency[:CW :IMMediate]?

Returned Parameters

<NR3>

Related Commands

FREQ:MODE FREQ:SLEW

FREQuency:MODE

This command determines how the output frequency is controlled during a triggered output transient. The choices are:

FIXed

The output frequency is unaffected by a triggered output transient.

STEP

The output frequency is programmed to the value set by FREQuency:TRIGgered when a

 

triggered transient occurs.

PULSe

The output frequency is changed to the value set by FREQuency:TRIGgered for a duration

 

determined by the pulse commands.

LIST

The output frequency is controlled by the frequency list when a triggered transient occurs.

Command Syntax

Parameters

*RST Value

Examples

Query Syntax

Returned Parameters

Related Commands

[SOURce:]FREQuency:MODE <mode> FIXed STEP PULSe LIST FIXed

FREQ:MODE FIX

[SOURce:]FREQuency:MODE?

<CRD>

FREQ FREQ:TRIG

65

Page 65
Image 65
Agilent Technologies 6812B, 6834B, 6814B, 6843A, 6813B, 6811B Source Subsystem Frequency, FREQuency:MODE, Subsystem Syntax