Agilent Technologies 6813B, 6834B TRIGger:SEQuence2:PHASe TRIGger:SYNCHronize:PHASe, Query Syntax

Models: 6811B 6813B 6834B 6843A 6814B

1 187
Download 187 pages 38.86 Kb
Page 111
Image 111

Language Dictionary - 3

TRIGger:SEQuence2:PHASe

TRIGger:SYNCHronize:PHASe

These commands set the phase angle with respect to an internal phase reference at which PHASe:SYNChronous:SOURce becomes true. The range is from 360 to +360 degrees.

Command Syntax

Parameters

*RST Value

Examples

Query Syntax

Returned Parameters Related Commands

TRIGger:SEQuence2:PHASe <NRf+>

TRIGger:SYNChronize:PHASe <NRf+>

360 to +360 (degrees) MAXimum MINimum

0

 

TRIG:SYNC:PHAS 90

TRIG:SEQ2:PHAS 180

TRIGger:SEQuence2:PHASe?

TRIGger:SYNChronize:PHASe?

<NR3>

ABOR TRIG TRIG:DEL TRIG:SYNC INIT INIT:CONT *TRG WAI

TRIGger:SEQuence3

TRIGger:ACQuire

When the trigger subsystem has been initiated, these commands generate a measurement trigger regardless of the selected trigger source. The measurement trigger causes the to digitize the instantaneous output voltge and current for several output cycles and store the results in a buffer.

The FETCh commands return the requested calculation from this acquired data. When the measurement completes, the WTG bit in the Status Operation Condition register is cleared.

Command Syntax

TRIGger:SEQuence3[:IMMediate]

 

TRIGger:ACQuire:[:IMMediate]

Parameters

None

 

 

Examples

TRIG:ACQ

TRIG:SEQ3:IMM

Related Commands

ABOR TRIG

TRIG:DEL

TRIG:SYNC

 

TRIG:SYNC:PHAS INIT

INIT:CONT *TRG WAI

111

Page 111
Image 111
Agilent Technologies 6813B TRIGger:SEQuence2:PHASe TRIGger:SYNCHronize:PHASe, TRIGger:SEQuence3 TRIGger:ACQuire, Examples