Agilent Technologies 6843A, 6834B, 6814B manual MEASure:POWer:AC:APParent?, FETCh:POWer:AC:APParent?

Models: 6811B 6813B 6834B 6843A 6814B

1 187
Download 187 pages 38.86 Kb
Page 50
Image 50

3 - Language Dictionary

MEASure:POWer:AC:APParent?

FETCh:POWer:AC:APParent?

Phase Selectable

These queries return the apparent power being sourced at the output terminals in volt-amperes.

Query Syntax

MEASure:[SCALar]:POWer:AC:APParent?

 

FETCh:[SCALar]:POWer:AC:APParent?

Parameters

None

 

Examples

MEAS:POW:AC:APP?

FETC:POW:AC:APP?

Returned Parameters

<NR3>

 

Related Commands

MEAS:POW:REAC?

MEAS:POW:PFAC?

MEASure:POWer:AC:REACtive?

FETCh:POWer:AC:REACtive?

Phase Selectable

These queries return the reactive power being sourced at the output terminals in volt-amperes reactive. Reactive power is computed as:

VAR = sqrt(square(apparent power) square(real power))

Query Syntax

MEASure:[SCALar]:POWer:AC:REACtive?

 

FETCh:[SCALar]:POWer:AC:REACtive?

Parameters

None

 

Examples

MEAS:POW:AC:REAC?

FETC:POW:AC:REAC?

Returned Parameters

<NR3>

 

Related Commands

MEAS:POW:AC:APP?

MEAS:POW:PFAC?

MEASure:POWer:AC:PFACtor?

FETCh:POWer:AC:PFACtor?

Phase Selectable

These queries return the output power factor. The power factor is computed as: pfactor = real power/apparent power

Query Syntax

MEASure:[SCALar]:POWer:AC:PFACtor?

 

FETCh:[SCALar]:POWer:AC:PFACtor?

Parameters

None

 

Examples

MEAS:POW:AC:PFAC?

FETC:POW:AC:PFAC?

Returned Parameters

<NR3>

 

Related Commands

MEAS:POW:AC:APP?

MEAS:POW:REAC?

50

Page 50
Image 50
Agilent Technologies 6843A, 6834B, 6814B MEASure:POWer:AC:APParent?, FETCh:POWer:AC:APParent?, MEASure:POWer:AC:REACtive?