Agilent Technologies 6834B ABORt, INITiate SEQuence INITiate NAME, Abor, INIT SEQ1, Init Name Acq

Models: 6811B 6813B 6834B 6843A 6814B

1 187
Download 187 pages 38.86 Kb
Page 108
Image 108
ABORt

3 - Language Dictionary

ABORt

This command resets the measurement and transient trigger systems to the Idle state. Any output transient or measurement that is in progress is immediately aborted. ABORt also cancels any lists or pulses that may be in process.

ABORt also resets the WTG bit in the Operation Condition Status register (see Chapter 4 under “Programming the Status Registers”). ABORt is executed at power turn-on and upon execution of *RCL, RST, or any implied abort command (see List Subsystem).

NOTE: If INITiate:CONTinuous ON has been programmed, the trigger subsystem initiates itself immediately after ABORt, thereby setting the WTG bit.

Command Syntax

ABORt

Parameters

None

Examples

ABOR

Related Commands

INIT *RST *TRG TRIG

INITiate:SEQuence

INITiate:NAME

The INITiate commands control the initiation of both the transient generator and the measurement trigger systems. They cause the trigger system to make a transition from the Idle state to the Waiting-for-Trigger state. If the trigger system is not in the Idle state, the initiate commands are ignored.

INITiate:SEQuence and INITiate:NAME initiate the trigger systems to reference trigger sequences. INITiate:SEQuence references a trigger sequence by its number, while INITiate:NAME references a sequence by its name. The correspondence between sequence names and numbers is:

Sequence Number

Sequence Number

Description

1 (the default)

TRANsient

Step, pulse, or list transient trigger sequence

3

ACQuire

Measurement acquire trigger sequence

Command Syntax

INITiate[:IMMediate]:SEQuence[ 1 3 ]

 

INITiate[:IMMediate]:NAME<name>

Parameters

For INIT:NAME: TRANsient ACQuire

Examples

INIT:SEQ1

INIT:NAME ACQ

Related Commands

ABOR INIT:CONT TRIG *TRG

108

Page 108
Image 108
Agilent Technologies 6834B, 6814B, 6843A, 6813B, 6811B ABORt, INITiate SEQuence INITiate NAME, Abor, INIT SEQ1, Init Name Acq