Agilent Technologies 6843A SYSTem:LANGuage, SYSTem:LOCal, SYSTem:REMote, SYSTem:RWLock, Syst:Loc

Models: 6811B 6813B 6834B 6843A 6814B

1 187
Download 187 pages 38.86 Kb
Page 104
Image 104

3 - Language Dictionary

SYSTem:LANGuage

Sets the command language of the ac source to either SCPI or Elgar Model 9012 PIP. The language selection is stored in non-volatile memory and is retained after power-off. Both the command and query form can be given regardless of the current language. Refer to Appendix D for more information.

Command Syntax SYSTem:LANGuage <language>

Parameters SCPI E9012

Example SYST:LANG SCPI

Query Syntax SYSTem:LANGuage?

Returned Parameters <CRD>

SYSTem:LOCal

This command can only be used with the RS-232 interface. It sets the interface in Local state, which enables the front panel controls.

Command Syntax

SYSTem:LOCal

Parameters

None

Example

SYST:LOC

Related Commands

SYST:REM SYST:RWL

SYSTem:REMote

This command can only be used with the RS-232 interface. It sets the interface in the Remote state, which disables all front panel controls except the Local key. Pressing the Local key while in the Remote state returns the front panel to the Local state.

Command Syntax

SYSTem:REMote

Parameters

None

Example

SYST:REM

Related Commands

SYST:LOC SYST:RWL

SYSTem:RWLock

This command can only be used with the RS-232 interface. It sets the interface in the Remote-Lockout state, which disables all front panel controls including the Local key. Use SYSTem:LOCal to return the front panel to the Local state.

Command Syntax

SYSTem:RWLock

Parameters

None

Example

SYST:RWL

Related Commands

SYST:REM SYST:LOC

104

Page 104
Image 104
Agilent Technologies 6843A manual SYSTem:LANGuage, SYSTem:LOCal, SYSTem:REMote, SYSTem:RWLock, Syst:Loc, Syst:Rem, Syst:Rwl