Agilent Technologies 6814B, 6834B, 6843A, 6813B, 6811B Source Subsystem Voltage, Subsystem Syntax

Models: 6811B 6813B 6834B 6843A 6814B

1 187
Download 187 pages 38.86 Kb
Page 85
Image 85
Source Subsystem (Voltage)

Language Dictionary - 3

Source Subsystem (Voltage)

This subsystem programs the output voltage of the ac source.

Subsystem Syntax

[SOURce:]

 

VOLTage

 

[:LEVel]

 

[:IMMediate]

 

[:AMPLitude] <n>

Sets the ac rms voltage amplitude

:TRIGgered

 

[:AMPLitude] <n>

Sets the transient voltage amplitude

:MODE <mode>

Sets the voltage mode (FIX STEP PULS LIST)

:OFFSet

 

[:IMMediate] <n>

Sets the dc offset voltage

:MODE <mode>

Sets the offset mode (FIX STEP PULS LIST)

:TRIGgered <n>

Sets the transient dc offset voltage

:SLEW

 

[:IMMediate] <n> INFinity

Sets the voltage slew rate

:MODE <mode>

Sets voltage slew mode (FIX STEP PULS LIST)

:TRIGgered <n> INFinity

Sets the transient voltage slew rate

:PROTection

 

[:LEVel] <n>

Sets the overvoltage protection threshold

:STATe <bool>

Sets the overvoltage protection state

:RANGe <n>

Sets the voltage range

:SENSe ALC

 

:DETector RTIMe RMS

Sets the sense detector for the voltage control loop

:SOURce INTernal EXTernal

Sets voltage sense source

:SLEW

 

[:IMMediate] <n> INFinity

Sets the voltage slew rate

:MODE <mode>

Sets voltage slew mode (FIX STEP PULS LIST)

:TRIGgered <n> INFinity

Sets the transient voltage slew rate

85