IEC Mode Command Summary - E

CALCulate:INTegral:TIME

This command selects the Pst integration time for IEC Flicker measurements. The parameter may only assume values of 1, 5, 10, and 15 minutes in accordance with IEC 868. The command will be accepted and may be queried, but will have no meaningful function unless the ac source is placed in IEC mode using the SYSTem:CONFigure command.

Command Syntax

CALCulate:INTegral:TIME <NRf+>

Parameters

1, 5, 10, & 15 minutes

*RST Value

10 minutes

Examples

CALC:INT:TIME 10

Query Syntax

CALCulate:INTegral:TIME?

Returned Parameters

<NR3>

Related Commands

SYSTEM:CONF MEAS:ARR:VOLT:FLUC:FLIC?

 

MEAS:ARR:VOLT:FLUC:PST?

 

MEAS:ARR:VOLT:FLUC:ALL?

CALCulate:SMOothing

This command turns on or off a smoothing filter for current harmonic measurements. The filter transfer function is equivalent to a single pole lowpass function with a 1.5 second time constant and is applied only to current harmonic measurements made when IEC mode is selected with SYSTem:CONF.

Command Syntax

CALCulate:SMOothing <Bool>

Parameters

0 1 ON OFF

*RST Value

OFF

Examples

CALC:SMO ON

Query Syntax

CALCulate:SMOothing?

Returned Parameters

<CRD>

Related Commands

MEAS:ARR:CURR:HARM? SYST:CONF

 

MEAS:CURR:HARM?