This subsystem programs the three-phase output capability of the Agilent 6834B .

Subsystem Syntax

INSTrument

 

:COUPle <phase>

Couple all phases for programming (ALL NONE)

:NSELect <n>

Select the output phase to program (1 2 3)

:SELect <output>

Select the output phase to program (OUTP1 OUTP2 OUTP3)

INSTrument:COUPle

Agilent 6834B Only

In a three-phase power source it is convenient to set parameters of all three output phases simultaneously with one programming command. When INST:COUP ALL is programmed, sending a command to any phase will result in that command being sent to all three phases.

NOTE: INSTrument:COUPle only affects the operation of subsequent commands. It does not by itself immediately affect the ’s output. The commands that are affected by INSTrument:COUPle are those with the designation: Phase Selectable.

INSTrument:COUPle has no affect on queries. There is no way to query more than one phase with a single command. Directing queries to individual phases is done with INSTrument:NSELect.

Command Syntax

INSTrument:COUPle <phase>

Parameters

ALL NONE

*RST Value

ALL

Examples

INST:COUP ALL

Query Syntax

INSTrument:COUPle?

Returned Parameters

<CRD>

Related Commands

INST:NSEL