Agilent Technologies 6843A, 6834B OUTPut:COUPling, OUTPut:DFI:SOURce, Outp:Coup Dc, Outp:Dfi

Models: 6811B 6813B 6834B 6843A 6814B

1 187
Download 187 pages 38.86 Kb
Page 56
Image 56

3 - Language Dictionary

OUTPut:COUPling

Agilent 6811B, 6812B, 6813B, Only

This command enables ac or dc output coupling. When the output coupling is set to AC, a dc leveling loop attempts to maintain zero average output voltage. The loop has a corner frequency of about 2Hz. It will not prevent short transient waveforms that may have non-zero average voltage, but will cause a settling transient to an average value of 0 volts.

The output coupling must be set to DC to obtain dc output with VOLTage:OFFSet, or to generate output transients that have net dc components.

Command Syntax

OUTPut:COUPling <coupling>

Parameters

AC DC

*RST Value

AC

Examples

OUTP:COUP DC

Query Syntax

OUTPut:COUPling?

Returned Parameters

<CRD>

Related Commands

*RCL *SAV

OUTPut:DFI

This command enables or disables the discrete fault indicator (DFI) signal to the ac source.

Command Syntax

OUTPut:DFI[:STATe] <bool>

Parameters

0 1 OFF ON

 

*RST Value

OFF

 

Examples

OUTP:DFI 1

OUTP:DFI ON

Query Syntax

OUTPut:DFI[:STATe]?

Returned Parameters

0 1

 

Related Commands

OUTP:DFI:SOUR

 

OUTPut:DFI:SOURce

This command selects the source for DFI events. The choices are:

QUEStionable

Questionable summary bit

OPERation

Operation summary bit

ESB

Standard Event summary bit

RQS

Request Service summary bit

OFF

Never true

Command Syntax

Parameters

*RST Value

Examples

Query Syntax

Returned Parameters

Related Commands

OUTPut:DFI:SOURce <source> QUEStionable OPERation ESP RQS OFF OFF

OUTP:DFI:SOUR OPER

OUTPut:DFI:SOURce?

<CRD>

OUTP:DFI

56

Page 56
Image 56
Agilent Technologies 6843A, 6834B, 6814B, 6813B, 6811B, 6812B OUTPut:COUPling, OUTPut:DFI:SOURce, Outp:Coup Dc, Outp:Dfi