Agilent Technologies 6814B LIST:VOLTage:OFFSet:SLEW:POINts?, LIST:VOLT:OFFSet:SLEW:POIN?

Models: 6811B 6813B 6834B 6843A 6814B

1 187
Download 187 pages 38.86 Kb
Page 79
Image 79

Language Dictionary - 3

LIST:VOLTage:OFFSet:SLEW

Agilent 6811B, 6812B, 6813B, Only

This command specifies the dc offset slew list points. The slew points are given in the command parameters, which are separated by commas. The order in which the points are entered determines the sequence in which they are output when a list is triggered. Changing list data while a subsystem is in list mode generates an implied ABORt.

Command Syntax

Parameters

Unit

Examples

Query Syntax

Returned Parameters

Related Commands

[SOURce:]LIST:VOLTage:OFFSet:SLEW <NRf+> INF {,<NRf+> INF} 0 to 9.9E37 MAXimum MINimum INFinity

V (volts per second)

LIST:VOLT:OFFS:SLEW 10,20,1E2

[SOURce:]LIST:VOLTage:OFFSet:SLEW? <NR3> {,<NR3>}

LIST:VOLT:SLEW:POIN? LIST:COUN LIST:DWEL LIST:STEP

LIST:VOLTage:OFFSet:SLEW:POINts?

Agilent 6811B, 6812B, 6813B, Only

This query returns the number of points specified in LIST:VOLTage:OFFSet:SLEW. Note that it returns only the total number of points, not the point values.

Query Syntax

[SOURce:]LIST:VOLTage:OFFSet:SLEW:POINTs?

Returned Parameters

<NR1>

Examples

LIST:VOLT:OFFSet:SLEW:POIN?

Related Commands

LIST:VOLT:OFFS