Agilent Technologies 6834B, 6814B manual STATus:OPERation:NTRansition, STATus OPERation PTRansition

Models: 6811B 6813B 6834B 6843A 6814B

1 187
Download 187 pages 38.86 Kb
Page 96
Image 96
STATus:OPERation:NTRansition

3 - Language Dictionary

STATus:OPERation:NTRansition

STATus:OPERation:PTRansition

These commands set or read the value of the Operation NTR (Negative-Transition) and PTR (Positive- Transition) registers. These registers serve as polarity filters between the Operation Enable and Operation Event registers to cause the following actions:

υWhen a bit in the Operation NTR register is set to 1, then a 1-to-0 transition of the corresponding bit in the Operation Condition register causes that bit in the Operation Event register to be set.

υWhen a bit of the Operation PTR register is set to 1, then a 0-to-1 transition of the corresponding bit in the Operation Condition register causes that bit in the Operation Event register to be set.

υIf the same bits in both NTR and PTR registers are set to 1, then any transition of that bit at the Operation Condition register sets the corresponding bit in the Operation Event register.

υIf the same bits in both NTR and PTR registers are set to 0, then no transition of that bit at the Operation Condition register can set the corresponding bit in the Operation Event register.

NOTE: Setting a bit in the PTR or NTR filter can of itself generate positive or negative events in the corresponding Operation Event register.

Command Syntax

Parameters

Default Value

Examples

Query Syntax

Returned Parameters Related Commands

STATus:OPERation:NTRansition <NRf+> STATus:OPERation:PTRansition <NRf+> 0 to 32767 MAXimum MINimum

0

 

STAT:OPER:NTR 32

STAT:OPER:PTR 1

STATus:OPERation:NTRansition?

STATus:OPERation:PTRansition? <NR1> (register value)

STAT:OPER:ENAB

96

Page 96
Image 96
Agilent Technologies 6834B, 6814B, 6843A, 6813B STATus:OPERation:NTRansition, STATus OPERation PTRansition, Query Syntax