3 - Language Dictionary

LIST:STEP

This command specifies how the list sequencing responds to triggers. The following parameters may be specified:

ONCE

AUTO

causes the list to advance only one point after each trigger. Triggers that arrive during a dwell delay are ignored

causes the entire list to be output sequentially after the starting trigger, paced by its dwell delays. As each dwell delay elapses, the next point is immediately output

Command Syntax

[SOURce:]LIST:STEP <step>

Parameters

ONCE AUTO

*RST Value

AUTO

Examples

LIST:STEP ONCE

Query Syntax

[SOURce:]LIST:STEP?

Returned Parameters

<CRD>

Related Commands

LIST:COUN LIST:DWEL

LIST:TTLTrg

This command sets the sequence of Trigger Out list points. Each point which is set ON will cause a pulse to be output at Trigger Out when that list step is reached. Those entries which are set OFF will not generate Trigger Out pulses.

The order in which the list points are given determines the sequence in which Trigger Out pulses will be output when a list is triggered. Changing list data while a subsystem is in list mode generates an implied ABORt.

Command Syntax

[SOURce:]LIST:TTLTrg <bool> {,<bool>}

Parameters

0 1 OFF ON

 

Examples

LIST:TTLT 1,0,1

LIST:TTLT ON,OFF,ON

Query Syntax

[SOURce:]LIST:TTLTrg?

Returned Parameters

0 1 {,0 1}

 

Related Commands

LIST:TTLT:POIN?

LIST:COUN LIST:DWEL LIST:STEP

OUTP:TTLT OUTP:TTLT:SOUR

LIST:TTLTrg:POINts?

This query returns the number of points specified in LIST:TTLT. Note that it returns only the total number of points, not the point values.

Query Syntax [SOURce:]LIST:TTLTrg:POINTs?

Returned Parameters <NR1>

Examples LIST:TTLT:POIN?

Related Commands LIST:TTLT