4 - Programming Examples

Step 4

Determine the number of times the list is repeated before it completes. For example, to

 

repeat a list 10 times use:

 

LIST:COUNt 10

 

Entering INFinity makes the list repeat indefinitely. At *RST, the count is set to 1.

Step 5

Determines how the list sequencing responds to triggers. For a closely controlled

 

sequence of output levels, you can use a dwell-paced list. To cause the list to be paced by

 

dwell time use:

 

LIST:STEP AUTO

 

As each dwell time elapses, the next point is immediately output. This is the *RST setting.

 

If you need the output to closely follow asynchronous events, then a trigger-paced list is

 

more appropriate. In a trigger-paced list, the list advances one point for each trigger

 

received. To enable trigger-paced lists use:

 

LIST:STEP ONCE

 

The dwell time associated with each point determines the minimum time that the output

 

remains at that point. If a trigger is received before the previous dwell time completes, the

 

trigger is ignored. Therefore, to ensure that no triggers are lost, program the dwell time to

 

zero.

Step 6

Use the transient trigger system to trigger the list. This is described in detail under

 

"Triggering Output Changes".

Triggering Output Changes

The ac source has two independent trigger systems. One is used for generating output changes, and the other is used for triggering measurements. This section describes the output trigger system. The measurement trigger system is described under "Triggering Measurements".

The basic components of both systems are the same, but the transient trigger system has additional delay and phase synchronization features that the measurement trigger system does not have. The following transient trigger sources can be selected:

IMMediate generates a trigger when the trigger system is initiated.

BUSselects GPIB bus triggers.

EXTernal selects the external Trigger In BNC connector.

SCPI Triggering Nomenclature

In SCPI terms, trigger systems are called sequences. When more than one trigger system exists, they are differentiated by naming them SEQuence1, SEQuence2, ... etc. In the ac source, SEQuence1 is the transient trigger system, SEQuence2 is the phase synchronization trigger system, and SEQuence3 is the measurement trigger system.

The ac source uses aliases with more descriptive names for these sequences. These aliases can be used instead of the sequence forms.

Sequence Form

Alias

SEQuence1

TRANsient

SEQuence2

SYNChronize

SEQuence3

ACQuire

132

Page 132
Image 132
Agilent Technologies 6834B Triggering Output Changes, Scpi Triggering Nomenclature, LISTCOUNt, Sequence Form Alias, 132

6812B, 6811B, 6813B, 6834B, 6843A specifications

Agilent Technologies, a leader in electronic test and measurement equipment, offers a range of powerful signal sources including the 6843A, 6834B, 6814B, 6813B, and 6811B models. These instruments are designed to support various applications in research, development, and manufacturing, providing precise signal generation capabilities.

The Agilent 6843A is a versatile signal generator known for its exceptional frequency range and modulation capabilities. It supports an extensive bandwidth, making it ideal for applications that require high-frequency signal generation. With its superior phase noise performance, the 6843A is an excellent choice for radar, wireless communications, and electronic warfare applications. The instrument features an intuitive user interface, allowing engineers to set parameters quickly and efficiently.

Next, the Agilent 6834B offers exceptional performance characteristics, including high output power and low distortion. This signal generator is particularly noted for its ability to produce complex modulation formats, making it suitable for testing advanced wireless communication systems. With a reliable and stable output, the 6834B ensures accurate and repeatable measurements, which is vital for thorough testing processes.

The 6814B model stands out for its dual-channel capabilities, allowing users to generate simultaneous signals for testing multiple components or systems. This feature significantly enhances testing efficiency and flexibility for engineers. With built-in arbitrary waveform functionality, users can create custom waveforms, making the 6814B suitable for a wide range of applications including device characterization and signal processing research.

For those seeking a more compact solution, the Agilent 6813B provides essential signal generation features without compromising on performance. It is designed for a variety of applications across telecommunications and consumer electronics, featuring a straightforward interface and robust performance metrics.

Lastly, the 6811B is an entry-level yet capable model that supports a broad spectrum of testing needs. Perfect for educational and laboratory environments, it provides essential functionalities required for effective signal generation and analysis.

Overall, Agilent Technologies' 6843A, 6834B, 6814B, 6813B, and 6811B signal generators offer an array of features and technologies that cater to various application needs. Their precision, reliability, and user-oriented designs position them as invaluable assets in any testing environment, ensuring engineers can carry out their work with confidence and accuracy.