Agilent Technologies 6813B, 6834B VOLTage SLEW MODE, VOLTage:SLEW:TRIGgered, Volt:Slew:Mode Step

Models: 6811B 6813B 6834B 6843A 6814B

1 187
Download 187 pages 38.86 Kb
Page 93
Image 93

Language Dictionary - 3

VOLTage:SLEW:MODE

Phase Selectable

This command determines how the output voltage slew rate is controlled during a triggered output transient. The choices are:

FIXed

The slew rate is unaffected by a triggered output transient.

STEP

The slew rate is programmed to the value set by VOLTage:SLEW:TRIGgered when a

 

triggered transient occurs.

PULSe

The slew rate is changed to the value set by VOLTage:SLEW:TRIGgered for a duration

 

determined by the pulse commands.

LIST

The slew rate is controlled by the voltage slew list when a triggered transient occurs.

Command Syntax

[SOURce:]VOLTage:SLEW:MODE <mode>

Parameters

FIXed STEP PULSe LIST

*RST Value

FIXed

Examples

VOLT:SLEW:MODE STEP

Query Syntax

[SOURce:]VOLTage:SLEW:MODE?

Returned Parameters

<CRD>

Related Commands

VOLT:SLEW VOLT:SLEW:TRIG

VOLTage:SLEW:TRIGgered

Phase Selectable

This command selects the slew rate that will be set during a triggered step or pulse transient. A parameter of MAXimum or INFinity sets the slew to its maximum possible rate. The SCPI representation for infinity is 9.9E37.

Command Syntax

[SOURce:]VOLTage:SLEW:TRIGgered <NRf+> INFinity

Parameters

0 to 9.9E37 MAXimum MINimum INFinity

Unit

V (volts per second)

 

*RST Value

INFinity

 

Examples

VOLT:SLEW:TRIG 50

VOLT:SLEW:TRIG MAX

Query Syntax

[SOURce:]VOLTage:SLEW:TRIGgered?

Returned Parameters

<NR3>

 

Related Commands

VOLT:SLEW VOLT:SLEW:MODE

93

Page 93
Image 93
Agilent Technologies 6813B, 6834B, 6814B, 6843A, 6811B, 6812B VOLTage SLEW MODE, VOLTage:SLEW:TRIGgered, Volt:Slew:Mode Step