Agilent Technologies 6812B, 6834B, 6814B, 6843A Signal, 1.Bit Configuration of Status Registers

Models: 6811B 6813B 6834B 6843A 6814B

1 187
Download 187 pages 38.86 Kb
Page 143
Image 143
Signal

NTR Filter

Event

Enable

STAT:OPER:NTR <n>

STAT:OPER:EVEN?

STAT:OPER:ENAB? <n>

Programming Examples - 4

A negative transition filter that functions as described under STAT:OPER:NTRPTR commands in Chapter 3. It is a read/write register.

A register that latches any condition that is passed through the PTR or NTR filters. It is a read-only register that is cleared when read.

A register that functions as a mask for enabling specific bits from the Event register. It is a read/write register.

The outputs of the Operation Status register group are logically-ORed into the OPER(ation) summary bit

(7) of the Status Byte register.

Bit

0

5

8

0

1

2

3

4

9

10

11

12

13

14

0

2

3

4

5

7

3

4

5

6

7

Signal

CAL

WTG

CV

OV

OCP

SOA

UNR

OT

RI

CLpeak

Rail

CLrms

Isum

MeasOvld

OPC

QYE

DDE

EXE

CME

PON

QUES

MAV

ESB

MSS

RQS

OPER

Table 4-1. Bit Configuration of Status Registers

Meaning

Operation Status Group

Interface is computing new cal constants

Interface is waiting for a trigger

The output voltage is regulated

Questionable and Questionable Instrument Isummary Status Groups

The overvoltage protection circuit has tripped The overcurrent protection circuit has tripped

The safe operating area protection has tripped (Agilent 6811B/6812B/6813B) The output is unregulated

An overtemperature condition has occurred The remote inhibit state is active

The peak current limit circuit is active (Agilent 6811B/6812B/6813B) The rail protection circuit has tripped (Agilent 6811B/6812B/6813B) The rail is unregulated (Agilent 6814B/6834B/6843A)

The rms current limit circuit is active

Summary of QUES:INST:ISUM registers (Agilent 6834B)

Current measurement exceeded capability of low range (Agilent 6811B/6812B/6813B)

Standard Event Status Group

Operation complete

Query error

Device-dependent error

Execution error

Command error

Power-on

Status Byte and Service Request Enable Registers

Questionable status summary bit

Message Available summary bit

Event Status Summary bit

Master Status Summary bit

Request Service bit

Operation status summary bit

143

Page 143
Image 143
Agilent Technologies 6812B, 6834B, 6814B Signal, 1.Bit Configuration of Status Registers, Meaning Operation Status Group