3 - Language Dictionary

MEASure:CURRent:CREStfactor?

FETCh:CURRent:CREStfactor?

Phase Selectable

These queries return the output current crest factor. This is the ratio of peak output current to rms output current.

Query Syntax

MEASure:[SCALar]:CURRent:CREStfactor?

 

FETCh:[SCALar]:CURRent:CRESfactor?

Parameters

None

 

Examples

MEAS:CURR:CRES?

FETC:CURR:CRES?

Returned Parameters

<NR3>

 

Related Commands

MEAS:CURR:ACDC?

MEAS:CURR:AMPL:MAX?

MEASure:CURRent:HARMonic?

FETCh:CURRent:HARMonic?

Phase Selectable

These queries return the rms amplitude of the Nth harmonic of output current.

The parameter is the desired harmonic number. Queries sent with a value of 0 return the dc component. A value of 1 returns the fundamental output frequency. Harmonic orders can be queried up to the fundamental measurement bandwidth of the measurement system, which is 12.6kHz. Thus the maximum harmonic that can be measured is dependent on the output frequency. Any harmonics that represent frequencies greater than 12.6kHz are returned as 0.

Query Syntax

MEASure:[SCALar]:CURRent:HARMonic[:AMPLitude]? <NRf>

 

FETCh:[SCALar]:CURRent:HARMonic[:AMPLitude]? <NRf>

Parameters

0 to 50

 

Examples

MEAS:CURR:HARM? 3

FETC:CURR:HARM? 1

Returned Parameters

<NR3>

 

Related Commands

MEAS:CURR:HARM:PHAS? MEAS:CURR:HARM:THD?

44