6-8 Calibration NetScan Users Manual
Gain Calibration of Main Unit
Required Equipment*
- Volts DC Calibrator
- 22-2 Twisted pairs (to connect calibrator to calibration card)
- Digital Voltmeter, or Digital Multimeter (used for verification of calibration voltages)
- Calibration Card (installed in bottom slot, slot #1, of main unit)
* see required specifications on page 6-2
&$87,21
Offset calibration of main unit must be completed prior to calibrating the main unit for
gain.
&$87,21
Use approved ESD precautions, including static-free work area and grounded wrist strap,
when handling circuit boards and electronic components. Failure to do so could cause
equipment damage due to electrostatic discharge.