4-38 ChartView Software Reference NetScan Users Manual
Acquisition Setup Dialog Box, High-Speed, Single-Channel Mode

Acquisition Setup

Parameter Options
Event Configuration Trigger: Set a trigger scan in the Acquis ition by using one of the following:
Keyboard - Use Console Trigger dialog box to start trigger scan.
Note: If the Keyboard Trigger is selected the use of Pre-trigger is disabled.
External TTL - Use an external signal to start the trigger scan.
Channel Value - Use specified value of a given channel.
Alarm - Use an Alarm value to start the trigger scan.
Absolute Time - Use a time of day to start the trigger scan.
Note: If Absolute Time is selected, the use of Auto Re-arm is disabled.
Stop: Set a stop scan in the Acquisition by using one of the above selections or Count (as
specified below). Keyboard isnt applicable. If a MultiScan, ChartScan, or NetSc an i s
connected and High-Speed Single-Channel (Burst Mode) is selected, then Count from
trigger is the only Stop available for the event configuration.
Count: Specify a number of scans to be read before a stop scan is set.
Acquisition Parameters Scan Counts: Specify the number of scans in an acquisi t i on
Pre-trigger
Post Stop
Average weight: Average weight can be specified only when in the Normal Mode. The value for
the average weight determines the number of samples to be averaged when in
Normal Mode. Values of 1, 2, 4, and 8 are for high-speed multiple channel
applications for thermocouples and DC voltages where noise filtering is not an
issue.
Values of 16, 32, 64, 128, and 256 are recommended for AC noise rejection.
Scan Intervals: Time between scans (frequency of scans)
Pre-trigger and Post Stop
Post-trigger
Use one Interval
High Speed Setup:
Allows for channel and scan rate entries for high-speed, single channel mode.
Note: These parameters cannot be altered while an acquisition is in progress.