NetScan Users Manual General Inform at ion and Specifications 3-5
Scan Interval:
Absolute time between scans (hh:mm:ss.t); min = 00:00:00.0,
max = 99:59:59.9. Note: Specifying a value of 00:00:00.0 results in no delay
between channel scans.
Triggers
Installation Category:
CE: Category 1.
Programmable Triggering:
Temperature or Voltage level (above or below), absolute time of day, alarm
condition (on or off), external TTL trigger (rising or falling), specified
number of readings.
Temperature-Level Trigger:
Programmable value for any one channel. Not available in Burst mode.
TTL Trigger:
Programmable for rising or falling edges.
Pre-Trigger Count and Post-Trigger Count:
Programmable integer (< memory size -1); and programmable integer..
Digital Filtering:
Average 16 samples at 50/60 Hz for line cycle noise.
TC Reference:
TC reference tables based on the IPTS-68 standard as listed with the US
Department of Commerce (NTIS). International Electrotechnical
Commission (IEC) 43 760.
Trigger Input Connector:
External BNC connector
Trigger Output Connector:
External BNC connector
Data Storage & Format
Storage:
128 K reading (256 Kbyte) standard; optional 500 K reading (1 Mbyte),
2 M reading (4 Mbyte), 4 M reading (8 Mbyte).
Data Formats:
ASCII and binary; binary format returns a 16-bit compensated and linearized
temperature value (0.1°C/bit); user-programmable for hi/low byte or low/hi
byte.
Statistical Parameters:
High, Low, and Last available per channel (during acquisition only). Not
available in single channel burst mode.
Time Stamp:
Available for each scan group and for each channels high, low, and last
parameters.
Time Format:
Relative time (hh:mm:ss.mil), absolute time (hh:mm:ss.mil) or absolute
date and time (dd/mm/yy hh:mm:ss.mil). Not available in single channel
burst mode.
Alarm Stamp:
Available for each scan group. Not available in single channel burst mode.