Appendix A API Command Reference
NetScan Users Manual A-27
H - Calibrate Channel Offset
TYPE
Calibration
EXECUTION
Immediate
SYNTAX
Hchan
Calibrate the correction offset constants for given option card where chan is
the channel number such that 1 < chan < max,
where max = 128 for NetScan. The chassis is chan = 0.
DESCRIPTION
Note: This command is only for advanced users who perform their own calibration. It i s not necessary for
normal, everyday operation.
Active only in calibration mode, the Calibrate Channel Offset (H) comm and is used to calibrate the correction
offset constants for a given slide-in card. The channel offset values are neces sary in order for the internal
firmware to accurately interpret readings coming from the input channels. This command will take the param eter
chan and calculate the offsets based on the present channel input voltage.
Note: To achieve correct results 0.0 mV must be present at the desired c hannel prior to performing the
command. The 0.0 mV value is obtained by placing a short across the applicabl e c hannel. Chapter 6,
Calibration, provides illustrations of various channel shorts.
After the command has been issued, the precision voltage source mus t remain at its value at least until the
trigger LED stops flashing in order to achieve accurate results. Refer to Chapter 6 f or m ore information
calibration.
EXAMPLE
PRINT#1, K12345X
Enter Calibration Mode
Connect short across applicable channel
PRINT#1, H17X
Calculate offsets for T/C card, card #2
Wait for command to complete
Remove precision voltage
PRINT#1, EX
End Calibration Mode