Main Menu (continued)
CALIBRATIONThe CALIBRATION function in this monitor measures LCD panel characteristics from low to high brightness values and uses internal monitor processing to handle calibration.
Calibration is performed using internal signals and do not rely on image quality settings. This monitor uses a color temperature of D65 for calibration and other color temperatures are derived from this value.
Equipment required for CALIBRATION•Konica Minolta
•Konica Minolta
•A
Connect the
•Set the Baud rate of the display color analyzer to 38,400.
This unit
Konica Minolta
Konica Minolta
| Analyzer Side | (Straight) | On the monitor | ||||||
| Pin | Signal |
|
|
|
|
| Pin | Signal |
| number |
|
|
|
|
| number | ||
|
|
|
|
|
|
| |||
1 | CD |
|
|
|
|
| 1 | N.C. | |
terminal |
|
|
|
|
| ||||
2 | RXD |
|
|
|
|
| 2 | TXD | |
(9 pins) |
|
|
|
|
| ||||
3 | TXD |
|
|
|
|
| 3 | RXD | |
|
|
|
|
|
| ||||
| 4 | DTR |
|
|
|
|
| 4 | DSR |
|
|
|
|
|
| ||||
| 5 | GND |
|
|
|
|
| 5 | GND |
|
|
|
|
|
| ||||
| 6 | DSR |
|
|
|
|
| 6 | DTR |
|
|
|
|
|
| ||||
| 7 | RTS |
|
|
|
|
| 7 | CTS |
|
|
|
|
|
| ||||
| 8 | CTS |
|
|
|
|
| 8 | RTS |
|
|
|
|
|
| ||||
| 9 | GND |
|
|
|
|
| 9 | N.C. |
•Turn on this monitor and perform adequate aging (about 1 hour) before starting calibration.
•Make the room dark so that no external light can enter the standard measurement probe before starting the calibration. When external light enters the probe, the low brightness characteristics may not be calibrated correctly.
•LCD panel characteristics and instrument error in the display color analyzer may sometimes result in small differences in values after calibration.
To further fine tune the monitor, set GAIN and BIAS for R, G and B in the VAR mode of [COLOR TEMP.].
•Do not apply the probe to a WFM/VECTOR display on the screen.
•When you are using
CALIBRATION
Select [CALIBRATION] in the [SYSTEM CONFIG] menu to open the following menus.
Submenu | Setting | Description |
AUTO CALIBRATION *8 | ▼ | Connect a |
|
| calibration. |
|
| Select [AUTO CALIBRATION] and select [YES] in the confirmation screen that |
|
| appears to start calibration. |
RESET *9 | ▼ | Returns calibration data to their factory defaults. |
|
| Select [RESET] and select [YES] in the confirmation screen that appears to return |
|
| calibrated values to their factory defaults. |
*8 [EXECUTING] is displayed during AUTO CALIBRATION and [COMPLETE] appears when calibration ends. [INCOMPLETE] appears if calibration could not be completed.
*9 When RESET ends, [COMPLETE] appears.