Main Menu  (continued)

Performing AUTO CALIBRATION

1.Turn the rotary knob <PICTURE> in the [SYSTEM CON- FIG] menu to select the [CALIBRATION] submenu and press the rotary knob <PICTURE>.

2.Turn the rotary knob <PICTURE> to select the [AUTO CALIBRATION] submenu and press the rotary knob <PICTURE>.

[AUTO CALIBRATION]

YES

NO

MENU EXIT

SEL.

ENTER RETURN RETURN

3.Select [YES] and press the rotary knob <PICTURE>. This starts calibration.

[AUTO CALIBRATION]

xxxxxxxxx

MENUEXIT

RETURN RETURN

xxxxxxxxxindicates that the following message with the meaning listed below appears.

EXECUTING:

Operation in progress

COMPLETE:

Operation completed

INCOMPLETE:

Operation incomplete (Check the

 

RS-232C connection or display color

 

analyzer connection.)

4.Turn the power off and then on after performing auto calibration.
RESET operation

1.Turn the rotary knob <PICTURE> in the [SYSTEM CON- FIG] menu to select the [CALIBRATION] submenu and press the rotary knob <PICTURE>.

2.Turn the rotary knob <PICTURE> to select the [RESET] submenu and press the rotary knob <PICTURE>.

[RESET]

YES

NO

MENU EXIT

SEL.

ENTER RETURN RETURN

3.Select [YES] and press the rotary knob <PICTURE>. This starts calibration.

[RESET]

xxxxxxxxx

MENUEXIT

RETURN RETURN

xxxxxxxxx indication.

COMPLETE: Operation completed

4.Turn the power off and then on after performing auto calibration.

33