C A L I B R A T I N G T H E V E R I F I E R

3

This chapter explains how to calibrate the verifier using the RJS ScanVision software.

The ScanVision software is found on the Tabletops Documentation

CD-ROM.Run the setup.exe file and follow the prompts as necessary.

C a l i b r a t i n g t h e V e r i f i e r

At initial setup, calibrate the verifier to ensure the scanning distance and orientation is within device limits. A calibration label is included in the verifier’s box. Store the calibration label in a clean location.

1.Remove all bar codes from the laser beam path.

2.Place the supplied calibration label in the laser beam in the same position (distance and angle) as the labels to be verified will be scanned.

3.Press and hold the RESET button on the verifier until the calibration LED begins to flash.

4.Release the RESET button immediately after the calibration LED begins flashing.

5.If calibration is successful, the laser beam turns off and the calibration LED turns off. Remove the calibration label from the beam path and Press the RESET button until the beam turns on.

The verifier is ready to operate.

6.If calibration is unsuccessful, the calibration LED is either on steadily or flashing. In this case, repeat the calibration procedure.

Once calibration is successful, re-calibration is not required unless the verifier unit has been moved – either on its mounting stand or to a new location. Keep the laser beam exit window clean. Dirt, dust, fingerprints, etc. on the exit window can affect calibration. See the cleaning procedures at the beginning of this manual keep the scanning window clean.

Calibrating the Verifier 3-1