Performance Tests: Electrical Safety Tests

Figure 6-5

Transducer Leakage Current Test Diagram

 

Power cord connected to AC power.

Hot -

Neutral -

(open neutral)

 

 

(open ground)

 

Ground wire:

 

Green or Green/Yellow

System under test

Powered on

Transducer submerged in saline

 

AC microammeter

Ground wire closed for Normal condition,

 

 

 

 

 

 

 

 

 

 

I

Ground wire open for first Single Fault condition,

 

 

 

 

 

 

 

(current)

Neutral wire open (ground closed) for second Single Fault condition

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ECG lead wire

LIMITS: UL, IEC, and CSA:

For

 

type transducers:

 

 

 

 

I = 10 microamperes Normal condition,

50 microamperes Single Fault condition

For type transducers:

I = 100 microamperes Normal condition, 500 microamperes Single Fault condition

7ASW027

Transducer Leakage Current Test Procedure

1.Set the analyzer mode to test ECG leads.

2.Plug the analyzer into an available AC wall outlet. Plug the ultrasound system power plug into the test receptacle on the analyzer.

3.Power up the ultrasound system by starting its internal PC.

4.Plug the transducer to be tested into the ultrasound system. Connect an ECG lead wire to the appropriate jack on the analyzer.

Page 100
Image 100
Philips M2540 service manual Transducer Leakage Current Test Diagram, Transducer Leakage Current Test Procedure

M2540 specifications

The Philips M2540 is a versatile and feature-rich patient monitor designed to enhance healthcare delivery in various clinical settings, from hospitals to outpatient facilities. With its advanced technology and user-friendly interface, the M2540 optimizes patient care through accurate monitoring and seamless data integration.

One of the standout features of the Philips M2540 is its advanced parameter capabilities. It supports a wide range of measurements including ECG, heart rate, non-invasive blood pressure, temperature, and respiratory rate. This comprehensive monitoring allows healthcare professionals to obtain a holistic view of a patient's condition in real time, ensuring timely interventions when necessary.

In addition to the standard parameters, the M2540 offers optional modules for more specialized monitoring, such as multi-gas analysis and invasive blood pressure. This adaptability makes it suitable for various clinical environments, including surgical units, critical care, and ambulatory settings.

The M2540 is equipped with Philips' innovative IntelliVue technology, which enhances monitoring accuracy and reliability. IntelliVue incorporates advanced algorithms that reduce false alarms, thus easing the burden on healthcare staff and improving patient safety. Furthermore, the monitor features integrated patient data management capabilities, enabling seamless data transfer to electronic health records (EHRs) and facilitating better communication among the care team.

User experience is prioritized in the design of the M2540, featuring a clear and intuitive display that presents vital signs in a straightforward manner. The touchscreen interface is customizable, allowing users to arrange critical data according to their preferences. This ensures that clinicians can access vital information quickly and efficiently during high-pressure situations.

The portability of the M2540 is another significant advantage. It is designed to be lightweight and easily transportable, making it ideal for use in various locations, from operating rooms to patient transport. The long-lasting battery life ensures uninterrupted monitoring, even during transfers.

In summary, the Philips M2540 is a robust patient monitoring solution that combines advanced technology with user-centric design. Its comprehensive monitoring capabilities, IntelliVue technology, and portability make it an asset in any healthcare environment, ultimately enhancing the quality of patient care and outcomes.