Philips M2540 service manual System Motherboard Distribution Board Connector Modules

Models: M2540

1 220
Download 220 pages 4.89 Kb
Page 44
Image 44

M2540-92000-01 A M2540 Ultrasound System Field Service ManualPage 44 Theory of Operation: System Motherboard

Primary

Functions

Connectors

System Motherboard

Distribution

Board and

Connector

Modules

The module’s primary functions are as follows:

Process electrocardiogram (ECG) signals from patient electrodes and deliver the data to the M2540A’s PC

Send a detected R-wave trigger to the APIO board for frame triggering in various scanning modes

Provide gain control for ECG waveforms

The Physio module has two connectors on its front panel:

The ECG-input connector is a round, 12-pin connector that is isolated from chassis ground.

A ¼-inch phone plug jack accepts auxiliary input and has both signal and shield grounds.

This board connects all data paths between the TR boards, the BPAP board, and the distribution board. It is also the distribution point for DC power and ground to all boards (the BPAP board and distribution board also receive power directly from the system power supply.). It also sends filtered high voltage to the TR boards.

The distribution board passes transducer signal inputs to the imaging system. The system power supply provides high voltage directly to the distribution board to fire transducers.

Signals that pass through this board include

Transducer identification signals from the BPAP board to and from the transducer

Transmit and receive signals from the TR boards to the active transducer

TEE motor-control and temperature-sense signals from the BPAP to and from the TEE trans- ducer

Page 44
Image 44
Philips M2540 service manual System Motherboard Distribution Board Connector Modules, Primary Functions Connectors