M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 56

Theory of Operation: Transducer Safety Testing: Test Setup and Theory

Figure 4-6 Transducer Leakage Current Test Diagram for 1st Single Fault ConditionWhen an open ground condition is imposed, all Ichassis is forced through the transducer. This Itransducer is still fairly low unless its sheath is compromised.Figure 4-7 Transducer Leakage Current Test Diagram for 2nd Single Fault Condition
A
I
7ASW030-3
Hot
Neutral
Ground
I
C
I
I
measured
system
chassis
transducer
S
Z
A
I
7ASW030-4
Hot
Neutral
Ground
I
C
I
I
measured
system
chassis
transducer
S
Z