M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 149
Service Procedures: PC Service
c. Slide the bezel to the right until it can be removed.
d. Each of the replaceable parts listed in step6 is retained by two plastic latches. To remove
the part, press the two latches in toward the part, and slide the part out of the PC case.
Reverse the procedure to reassemble the PC. Perform all RST and system checks to verify
performance.