M2540-92000-01 A M2540 Ultrasound System Field Service ManualPage 156 Service Procedures: Transducer Connector Modules

Transducer

Connector

Modules

The screws that secure the transducer connector modules are behind the service panel on the front of the lower enclosure.

Replacing Transducer Connector Modules

To replace a transducer connector module, see Figure 13-2on page 183 to locate parts and assemblies, and complete the following procedure:

1.Disconnect the system power cord from the main AC supply.

2.Remove the service panel at the right front of the lower enclosure by pressing down the two latches at the panel’s top.

3.Remove the screws that secure the module bezel to the e-box, remove the bezel, and pull the module out.

CAUTION

Be careful when removing and installing the transducer connector modules, to prevent damage

 

to the distribution board’s connectors.

 

 

 

 

4.

Reverse step 3 through step 1 to install a transducer connector module.

 

5.

Perform all RST and system checks to verify performance.

Fan

The M2540A’s cooling fan is attached to the front of the power supply. Replacement fans are

 

supplied with the finger guard installed.

 

Replacing the Fan

 

To remove the fan, see Figure 13-2on page 183 to locate parts and assemblies, and complete the

 

following procedure:

 

1.

Disconnect the system power cord from the main AC supply.

Page 156
Image 156
Philips M2540 service manual Replacing Transducer Connector Modules, Replacing the Fan

M2540 specifications

The Philips M2540 is a versatile and feature-rich patient monitor designed to enhance healthcare delivery in various clinical settings, from hospitals to outpatient facilities. With its advanced technology and user-friendly interface, the M2540 optimizes patient care through accurate monitoring and seamless data integration.

One of the standout features of the Philips M2540 is its advanced parameter capabilities. It supports a wide range of measurements including ECG, heart rate, non-invasive blood pressure, temperature, and respiratory rate. This comprehensive monitoring allows healthcare professionals to obtain a holistic view of a patient's condition in real time, ensuring timely interventions when necessary.

In addition to the standard parameters, the M2540 offers optional modules for more specialized monitoring, such as multi-gas analysis and invasive blood pressure. This adaptability makes it suitable for various clinical environments, including surgical units, critical care, and ambulatory settings.

The M2540 is equipped with Philips' innovative IntelliVue technology, which enhances monitoring accuracy and reliability. IntelliVue incorporates advanced algorithms that reduce false alarms, thus easing the burden on healthcare staff and improving patient safety. Furthermore, the monitor features integrated patient data management capabilities, enabling seamless data transfer to electronic health records (EHRs) and facilitating better communication among the care team.

User experience is prioritized in the design of the M2540, featuring a clear and intuitive display that presents vital signs in a straightforward manner. The touchscreen interface is customizable, allowing users to arrange critical data according to their preferences. This ensures that clinicians can access vital information quickly and efficiently during high-pressure situations.

The portability of the M2540 is another significant advantage. It is designed to be lightweight and easily transportable, making it ideal for use in various locations, from operating rooms to patient transport. The long-lasting battery life ensures uninterrupted monitoring, even during transfers.

In summary, the Philips M2540 is a robust patient monitoring solution that combines advanced technology with user-centric design. Its comprehensive monitoring capabilities, IntelliVue technology, and portability make it an asset in any healthcare environment, ultimately enhancing the quality of patient care and outcomes.