M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 150
Service Procedures: PC Service
Removing the PC from the System
To remove the PC, see Figure13-3 on page 185 to locate parts and assemblies, and complete the
following procedure:
1. Disconnect the system power cord from the main AC supply.
2. Remove the three screws from the lower edge of the lower enclosure’s left side panel, pull
the bottom of the panel out, and lower the panel clear of the system.
3. Disconnect all the cables from the PC. Note their locations so you can reinstall them
properly.
4. Loosen the screws in the angle bracket on the wall toward the center of the system. Remove
the screws that secures the PC to the angle bracket.
5. Remove the PC through the side of the system.
To install the PC in the system, reverse the procedure above. Perform all RST and system checks
to verify performance.