M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 129
Adjustments: Monitor Adjustment
6. Use the two right-hand monitor buttons to scroll up or down in the OSD Main menu and
select SIZE & POSITION.
7. Press the far left monitor button.
The Size & Position menu displays.
8. Use the two right-hand monitor buttons to select VERTICAL SIZE.
9. Press the far left monitor button.
The vertical size adjustment bar displays.
10. Use the two right-hand monitor buttons to adjust the vertical size so the bottom of the
display is visible.
11. Press the far left monitor button.
The vertical size adjustment bar closes.
12. Use the two right-hand monitor buttons to select SAVE & RETURN in the SIze & Position
menu.
13. Press the far left monitor button.
The SIze & Position menu closes.
14. In the OSD Main menu, use the two right-hand monitor buttons to select EXIT OSD.