M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 162
Service Procedures: Cables
6. Remove the four P-clips from the speaker supports.
7. Open the two service doors at the rear of the lower enclosure by turning their latches
counter-clockwise.
8. Disconnect the keyboard auxiliary power cable, the monitor power cable, and the control
panel ground cable from the power supply.
9. Disconnect the keyboard USB cable, the keyboard audio line cables, the microphone cable,
and the VGA cable from the PC.
10. Feed the lower ends of the cables into the lower enclosure central plenum, while pulling on
the upper end of the cable bundle.
11. Reverse step 10 through step 1 to install the cable bundle.
12. Perform all RST and system checks to verify performance.