M2540-92000-01 A M2540 Ultrasound System Field Service Manual

Page 59

 

Installation: Overview

5 Installation

Overview

This chapter includes procedures for installing and setting up the M2540A. The following sections are in this chapter:

“Assemble the System” on page 59

“System Startup” on page 62

“System Configuration” on page 64

“Customer Training” on page 88

Assemble the System

Power Cord and

Monitor

Installation

Peripherals

Installation

Complete the following steps to assemble the M2540A ultrasound system:

Installing the Power Cord and Monitor

1.Attach the system power cord to the receptacle at the bottom rear of the system.

2.Loosen the screw that secures the cord retainer, rotate the retainer 180 degrees, and tighten the screw.

3.Rotate the twivel base to its proper position. (The monitor cables pass through the rear of the twivel base.)

4.Install the monitor on the twivel.

5.Connect the VGA cable the and monitor power cable to the monitor.

When installing peripherals, keep the following points in mind:

No more than two peripherals can be mounted on the cart.

When two peripherals are installed, the smaller one must be on top of the larger one.