M2540-92000-01 A M2540 Ultrasound System Field Service Manual

Page 141

 

Service Procedures: Record Keeping

Rear Service Doors

Two doors in the rear of the lower enclosure give access to the rear of the PC, the e-box, and the power supply. The doors are hinged near the center of the system, and are held closed by quarter-turn latches near the sides of the system. To open the doors, turn the latches counter-clockwise.

NOTE References to the left and right sides of the system are as viewed from the front of the system.

Front Service Panel

Side Panels

Record

Keeping

A panel at the right side of the system’s front gives access to the transducer connector module mounting, the air filter, the fan, the front of the e-box, and the front of the power supply. To open this panel, press down the two latches at its top, pull the top edge out, and lift the panel clear of the system.

The two lower enclosure side panels are attached with screws on their lower edge. To remove either panel, remove the three screws along its lower edge, pull the bottom of the panel out, and lower the panel clear of the system.

You can make notes in the RST application’s service log, to help with later service procedures. To read service log entries, see “Using the Service Log” on page 125. To make an entry in the RST service log, see “Making an Entry in the Service Log” on page 125.

Page 141
Image 141
Philips M2540 service manual Record Keeping, Side Panels

M2540 specifications

The Philips M2540 is a versatile and feature-rich patient monitor designed to enhance healthcare delivery in various clinical settings, from hospitals to outpatient facilities. With its advanced technology and user-friendly interface, the M2540 optimizes patient care through accurate monitoring and seamless data integration.

One of the standout features of the Philips M2540 is its advanced parameter capabilities. It supports a wide range of measurements including ECG, heart rate, non-invasive blood pressure, temperature, and respiratory rate. This comprehensive monitoring allows healthcare professionals to obtain a holistic view of a patient's condition in real time, ensuring timely interventions when necessary.

In addition to the standard parameters, the M2540 offers optional modules for more specialized monitoring, such as multi-gas analysis and invasive blood pressure. This adaptability makes it suitable for various clinical environments, including surgical units, critical care, and ambulatory settings.

The M2540 is equipped with Philips' innovative IntelliVue technology, which enhances monitoring accuracy and reliability. IntelliVue incorporates advanced algorithms that reduce false alarms, thus easing the burden on healthcare staff and improving patient safety. Furthermore, the monitor features integrated patient data management capabilities, enabling seamless data transfer to electronic health records (EHRs) and facilitating better communication among the care team.

User experience is prioritized in the design of the M2540, featuring a clear and intuitive display that presents vital signs in a straightforward manner. The touchscreen interface is customizable, allowing users to arrange critical data according to their preferences. This ensures that clinicians can access vital information quickly and efficiently during high-pressure situations.

The portability of the M2540 is another significant advantage. It is designed to be lightweight and easily transportable, making it ideal for use in various locations, from operating rooms to patient transport. The long-lasting battery life ensures uninterrupted monitoring, even during transfers.

In summary, the Philips M2540 is a robust patient monitoring solution that combines advanced technology with user-centric design. Its comprehensive monitoring capabilities, IntelliVue technology, and portability make it an asset in any healthcare environment, ultimately enhancing the quality of patient care and outcomes.