M2540-92000-01 A M2540 Ultrasound System Field Service Manual Page 141
Service Procedures: Record Keeping
Rear Service
Doors
Two doors in the rear of the lower enclosure give access to the rear of the PC, the e-box, and
the power supply. The doors are hinged near the center of the system, and are held closed by
quarter-turn latches near the sides of the system. To open the doors, turn the latches
counter-clockwise.
NOTE References to the left and right sides of the system are as viewed from the front of the
system.
Front Serv ice
Panel
A panel at the right side of the system’s front gives access to the transducer connector module
mounting, the air filter, the fan, the front of the e-box, and the front of the power supply. To
open this panel, press down the two latches at its top, pull the top edge out, and lift the panel
clear of the system.
Side Panels
The two lower enclosure side panels are attached with screws on their lower edge. To remove
either panel, remove the three screws along its lower edge, pull the bottom of the panel out, and
lower the panel clear of the system.
Record Keep ing
You can make notes in the RST application’s service log, to help with later service procedures. To
read service log entries, see “Using the Service Log” on page 125. To make an entry in the RST
service log, see “Making an Entry in the Service Log” on page 125.