M2540-92000-01 A M2540 Ultrasound System Field Service Manual

Page 122

 

Performance Tests: Resident Self Tests

RST Log Codes

The Application Test Log’s Category column contains coded information. The code is in the following format:

SSnnNN

Where:

SS is the subsystem tested.

BP is the beam processor subsystem.

DP is the display processor subsystem.

nn is the module number of the tested module.

NN is the submodule number. Example: DP0102

The example indicates that the A/D converter (submodule 02 of module 01) of the display processor was tested.

To find the meaning of the numeric codes, use the tree control in the RST window’s left side as described in “Test Modules” on page 117.

Clearing the RST Test Log

1.Right-click RST Test Log in the tree view.

2.In the pop-up menu, select Clear Log.

3.Click Yes in the confirmation dialog box.