Teledyne 4060 manual Main Menus

Page 49

Series 4060

Operation

 

 

 

2.The MAIN MENU and SUB MENU screens have a flashing arrow on the left side indicating which item is being pointed to.

3.MODIFIABLE ITEM screens have a flashing value somewhere in the line.

4.6.4.1MAIN MENUS

The MAIN MENU consists of 15 functions you can use to customize and check the operation of the analyzer. They are listed here with brief descriptions:

SPAN: Set up and/or start a span calibration

ALT-SPAN: Set up and/or start an alternate span calibration.

MODEL: Displays Manufacturer, Model, and Software version of the instrument.

SELF-TEST:The instrument performs a self-diagnostic routine to check the integrity of the power supply, output boards, cell and amplifiers.

ALARMS: Used to set the alarm setpoints and determine whether each alarm will be active or defeated, HI or LO acting, and failsafe or not.

RANGE: Used to set up three analysis ranges that can be switched manually.

ANALOG-OUT ADJUST: Adjust the analog output with offset and scaling.

LINEARIZATION: Linearization option.

CHANGE STREAM: (optional feature) Controls a relay switch for using alternate gases. Has no effect during span mode (span gas relay setting used at that time).

AUTO-CAL:Used to define and/or start an automatic calibration sequence. AUTOCAL is an optional feature not included in the standard configurations of the 4060.

TIMING: Low level application specific timing parameters. Factory set. Do not change these.

Teledyne Analytical Instruments

49

Image 49
Contents Series Teledyne Analytical Instruments SeriesTeledyne Analytical Instruments Iii Safety MessagesSeries Combustible GAS Usage Additional Safety InformationSeries Teledyne Analytical Instruments Vii Procedure for Removal of Internal Inaccessible Shock HazardsViii Table of Contents Operation Maintenance & Troubleshooting Appendix B AppendixTeledyne Analytical Instruments Xiii List of FiguresList of Tables Introduction Main Features of the AnalyzerSeries Introduction Analyzer Description Principle of OperationIntroduction Series Model 4060 Front Panel Interface Applications Introduction Operational TheoryModes of Operation Series Operational TheoryOperational Theory Series Analyzer SubsystemsSample System Gas Flow Control System Internal Temperature ControllersFlame Ionization Detection Cell Fuel and Blanket Air SystemsTypical Piping Diagram for the Model Detection Cell Flame Ionization Cell Electrometer-AmplifierFlame Guard Circuit Anode Power SupplyFlame Ignition Circuit Series Installation InstallationUnpacking the Analyzer Mounting the AnalyzerElectrical Power Connections Installation SeriesUser Connections Electronic ConnectionsModel 4060 Rear Panel with Optional Gas Manifold Equipment Interface Connector Pin Arrangement Pin Function Threshold Alarm System Alarm Remote Calibration Connections Range ID Relay Connections 2.10 RS-232 Port Pin out of 50 pin D-Sub ConnectorGas Connections Commands via RS-232 InputGas Connections Less than 0.5 PPM Placing the System in Operation Equipment OperationSeries Operation Activating the Support Gases Preliminary Power-Off Check ListOperation Series 1 Air Flame IgnitionCarrier Gas Span GasAnalyzer Operation Sample PumpVerification of the Flame Guard Circuit Ignition and/or Flame Guard Circuit FailureSeries Operation Style Conventions Default ParametersEnter Navigation and Data EntryMenu Structure EscapeMain Menus Span ALT-SPAN System Self-Diagnostic TestModel Screen Alarms Function AL-1 Defeated AL-1 High AL-1 NON-FAILSAFE AL-1 NON-LATCHING Operation Series Range Function Analog Output AdjustmentLinearization Setting up an AUTO-CALChanging Stream Autocal Password Password Protection Group SetupTiming Entering a Password Enter NEW Password Screen Logging Out Repeat Password Entry ScreenStandby Series Operation Operation Series Teledyne Analytical Instruments Series Maintenance Maintenance & TroubleshootingMeasuring Circuit Electrical Checks Maintenance SeriesAnode Voltage Check Electronic Stability Printed Circuit Board ReplacementCollector Cable Temperature Control Electronic CheckIgnition and/or Flame Guard Circuit Checks Sampling System Printed Circuit Board Descriptions Flame Guard and Anode Power Supply PCBProportional Temperature Controller PCB Electrometer-Amplifier PCB Maintenance Series Appendix Specifications and Initial SettingsSeries Appendix Appendix Series Recommended Spare Parts ListAnalytical Instruments PC Board Assemblies Drawing ListAppendix Series Teledyne Analytical Instruments Appendix B B1 Addendum and Testing Results