Teledyne 4060 manual Pin out of 50 pin D-Sub Connector, 2.10 RS-232 Port

Page 36

Installation

Series 4060

 

 

 

 

 

 

Table 3-7: Pin out of 50 pin D-Sub Connector

pin #

Description

 

pin #

Description

1

 

 

26

 

2

 

 

27

 

3

+ Output 4-20 ma - Channel 2

 

28

Alarm 1 C Contact

4

- Output 4-20 ma - Channel 2

 

29

 

5

+ Output 4-20 ma – Channel 1

 

30

 

6

- Output 4-20 ma – Channel 1

 

31

 

7

- Output 0-1 v (Channel 1)

 

32

Exhaust Solenoid Hot

8

+ Output 0-1 v (Channel 2)

 

33

Sample Solenoid Hot

9

 

 

34

Range 4 Contact/ not used

10

Remote Span +

 

35

Range 4 Contact/not used

11

 

 

36

Alarm 3 NC Contact

12

Remote Span -

 

37

Alarm 3 NO Contact

13

 

 

38

Range 1 Contact

14

 

 

39

Range 2 Contact

15

 

 

40

Calibration Contact

16

Span Solenoid Return

 

41

Calibration Contact

17

Span Solenoid Hot

 

42

Alarm 2 NC Contact

18

Range 3 Contact

 

43

Alarm 2 NO Contact

19

Range 3 Contact

 

44

Alarm 2 C Contact

20

Alarm 3 C Contact

 

45

Alarm 1 NC Contact

21

Range 1 Contact

 

46

Alarm 1 NO Contact

22

Range 2 Contact

 

47

 

23

- Output 0-1 v (Channel 2)

 

48

Exhaust Solenoid Return

24

+ Output 0-1 v (Channel 1)

 

49

 

25

 

 

50

Sample Solenoid Return

3.3.2.10 RS-232 PORT

The digital signal output is a standard RS-232 serial communications port used to connect the analyzer to a computer, terminal, or other digital device. It requires a standard 9-pin D connector.

Input: The input functions using RS-232 that have been implemented to date are described in Table 3-8.

36

Teledyne Analytical Instruments

Image 36
Contents Series Series Teledyne Analytical InstrumentsSafety Messages Teledyne Analytical Instruments IiiSeries Additional Safety Information Combustible GAS UsageSeries Procedure for Removal of Internal Inaccessible Shock Hazards Teledyne Analytical Instruments ViiViii Table of Contents Operation Maintenance & Troubleshooting Appendix Appendix BList of Figures Teledyne Analytical Instruments XiiiList of Tables Main Features of the Analyzer IntroductionSeries Introduction Principle of Operation Analyzer DescriptionIntroduction Series Model 4060 Front Panel Interface Applications Operational Theory IntroductionModes of Operation Series Operational TheoryAnalyzer Subsystems Operational Theory SeriesSample System Internal Temperature Controllers Gas Flow Control SystemFuel and Blanket Air Systems Flame Ionization Detection CellTypical Piping Diagram for the Model Detection Cell Electrometer-Amplifier Flame Ionization CellAnode Power Supply Flame Guard CircuitFlame Ignition Circuit Installation Series InstallationUnpacking the Analyzer Mounting the AnalyzerInstallation Series Electrical Power ConnectionsUser Connections Electronic ConnectionsModel 4060 Rear Panel with Optional Gas Manifold Equipment Interface Connector Pin Arrangement Pin Function Threshold Alarm System Alarm Remote Calibration Connections Range ID Relay Connections Pin out of 50 pin D-Sub Connector 2.10 RS-232 PortCommands via RS-232 Input Gas ConnectionsGas Connections Less than 0.5 PPM Placing the System in Operation Operation EquipmentSeries Operation Preliminary Power-Off Check List Activating the Support GasesOperation Series Flame Ignition 1 AirCarrier Gas Span GasSample Pump Analyzer OperationVerification of the Flame Guard Circuit Ignition and/or Flame Guard Circuit FailureSeries Operation Default Parameters Style ConventionsNavigation and Data Entry EnterEscape Menu StructureMain Menus Span System Self-Diagnostic Test ALT-SPANModel Screen Alarms Function AL-1 Defeated AL-1 High AL-1 NON-FAILSAFE AL-1 NON-LATCHING Operation Series Analog Output Adjustment Range FunctionSetting up an AUTO-CAL LinearizationChanging Stream Autocal Password Group Setup Password ProtectionTiming Entering a Password Enter NEW Password Screen Repeat Password Entry Screen Logging OutStandby Series Operation Operation Series Teledyne Analytical Instruments Maintenance & Troubleshooting Series MaintenanceMaintenance Series Measuring Circuit Electrical ChecksAnode Voltage Check Printed Circuit Board Replacement Electronic StabilityTemperature Control Electronic Check Collector CableIgnition and/or Flame Guard Circuit Checks Sampling System Flame Guard and Anode Power Supply PCB Printed Circuit Board DescriptionsProportional Temperature Controller PCB Electrometer-Amplifier PCB Maintenance Series Specifications and Initial Settings AppendixSeries Appendix Recommended Spare Parts List Appendix SeriesAnalytical Instruments Drawing List PC Board AssembliesAppendix Series Teledyne Analytical Instruments B1 Addendum and Testing Results Appendix B