Installation | Series 4060 | |
|
|
|
|
|
|
3.3.2.6DIGITAL REMOTE CAL INPUTS
The digital remote calibration input accepts 0 V (off) or 24 VDC (on) for remote control of calibration. (See Remote Calibration Protocol below.) See Table
Span: Floating input. A
Cal Contact: This relay contact is closed while analyzer is spanning. (See Remote Calibration Protocol below.)
Table 3-5: Remote Calibration Connections
Pin Function
10+ Remote Span
12– Remote Span
40Cal Contact
41Cal Contact
Remote Calibration Protocol: To properly time the Digital Remote Cal Inputs to the Model 4060 Analyzer, the customer's controller must monitor the Cal Relay Contact.
When the contact is OPEN, the analyzer is analyzing, the Remote Cal Inputs are being polled, and a span command can be sent.
When the contact is CLOSED, the analyzer is already calibrating. It will ignore your request to calibrate, and it will not remember that request.
Once a span command is sent, and acknowledged (contact closes), release it. If the command is continued until after the span is complete, the calibration will repeat and the Cal Relay Contact (CRC) will close again.
For example:
1.Test the CRC. When the CRC is open, send a span command until the CRC closes (The CRC will quickly close.)
34 | Teledyne Analytical Instruments |