Page 50

Operation

Series 4060

 

 

 

 

GROUP SETUP: Determines compound names, associated timing and measurement ranges. Do not change these.

PSWD: Used to establish password protection or change the existing password.

LOGOUT: Logging out prevents unauthorized tampering with the analyzer settings.

STANDBY: Removes power to outputs and displays, but maintains power to internal circuitry.

Any function can be selected at anytime. Just scroll through the MAIN MENU with the DOWN/UP keys to the appropriate function, and ENTER it. The analyzer will immediately start that function, unless password restrictions have been assigned. (Password assignment is explained in Section 4.6.17).

4.6.5 Span

The SPAN function is used to calibrate the analyzer. Span can be performed either manually or automatically with the autocal function if equipped.

The analyzer is calibrated using span gas as described in Section 4.1. This section assumes that this gas has been properly connected and the line checked for leaks.

To initiate a span calibration:

<UP/DOWN >From the MAIN MENU, scroll down to the SPAN function.

<Enter> Press <Enter> to activate the SPAN function.

________________________________________________

-> Span 80.00 ppb Span Begin

________________________________________________

The first line allows the user to modify the span target value. The second line commences the span after the following query:

50

Teledyne Analytical Instruments

Image 50
Contents Series Series Teledyne Analytical InstrumentsSafety Messages Teledyne Analytical Instruments IiiSeries Additional Safety Information Combustible GAS UsageSeries Procedure for Removal of Internal Inaccessible Shock Hazards Teledyne Analytical Instruments ViiViii Table of Contents Operation Maintenance & Troubleshooting Appendix Appendix BList of Figures Teledyne Analytical Instruments XiiiList of Tables Series Introduction Main Features of the AnalyzerIntroduction Introduction Series Principle of OperationAnalyzer Description Model 4060 Front Panel Interface Applications Modes of Operation Operational TheoryIntroduction Series Operational TheorySample System Analyzer SubsystemsOperational Theory Series Internal Temperature Controllers Gas Flow Control SystemFuel and Blanket Air Systems Flame Ionization Detection CellTypical Piping Diagram for the Model Detection Cell Electrometer-Amplifier Flame Ionization CellFlame Ignition Circuit Anode Power SupplyFlame Guard Circuit Unpacking the Analyzer InstallationSeries Installation Mounting the AnalyzerUser Connections Installation SeriesElectrical Power Connections Electronic ConnectionsModel 4060 Rear Panel with Optional Gas Manifold Equipment Interface Connector Pin Arrangement Pin Function Threshold Alarm System Alarm Remote Calibration Connections Range ID Relay Connections Pin out of 50 pin D-Sub Connector 2.10 RS-232 PortCommands via RS-232 Input Gas ConnectionsGas Connections Less than 0.5 PPM Placing the System in Operation Series Operation OperationEquipment Operation Series Preliminary Power-Off Check ListActivating the Support Gases Carrier Gas Flame Ignition1 Air Span GasVerification of the Flame Guard Circuit Sample PumpAnalyzer Operation Ignition and/or Flame Guard Circuit FailureSeries Operation Default Parameters Style ConventionsNavigation and Data Entry EnterEscape Menu StructureMain Menus Span Model Screen System Self-Diagnostic TestALT-SPAN Alarms Function AL-1 Defeated AL-1 High AL-1 NON-FAILSAFE AL-1 NON-LATCHING Operation Series Analog Output Adjustment Range FunctionChanging Stream Setting up an AUTO-CALLinearization Autocal Password Timing Group SetupPassword Protection Entering a Password Enter NEW Password Screen Repeat Password Entry Screen Logging OutStandby Series Operation Operation Series Teledyne Analytical Instruments Maintenance & Troubleshooting Series MaintenanceAnode Voltage Check Maintenance SeriesMeasuring Circuit Electrical Checks Printed Circuit Board Replacement Electronic StabilityTemperature Control Electronic Check Collector CableIgnition and/or Flame Guard Circuit Checks Sampling System Flame Guard and Anode Power Supply PCB Printed Circuit Board DescriptionsProportional Temperature Controller PCB Electrometer-Amplifier PCB Maintenance Series Series Appendix Specifications and Initial SettingsAppendix Analytical Instruments Recommended Spare Parts ListAppendix Series Drawing List PC Board AssembliesAppendix Series Teledyne Analytical Instruments B1 Addendum and Testing Results Appendix B