Teledyne 4060 manual AL-1 Defeated AL-1 High AL-1 NON-FAILSAFE AL-1 NON-LATCHING

Page 53

Series 4060

Operation

 

 

 

-> AL-1 DEFEATED AL-1 HIGH

AL-1 NON-FAILSAFE

AL-1 NON-LATCHING

AL-1 20.00 ppb

AL-2 ACTIVE

AL-2 LOW

AL-2 FAILSAFE

AL-2 LATCHING

AL-2 50.00 ppb

________________________________________________

Note: For the purpose of example, Alarm 2 has been set differently than Alarm 1.)

Note: An alarm in 'alarm condition' is signaling that action must be taken such as correcting the alarm or the analysis concentration. If an active (not defeated) alarm has been set to HIGH at 20.0 ppb, and the analysis concentration is above that level, then the 'alarm condition' is occurring.

DEFEATED: If an alarm is defeated, its relay is de-energized, regardless of failsafe condition. A defeated alarm does not react to a transition over its trip point in either direction.

HIGH: If an alarm is set as HIGH, it will not create a new alarm condition (see latching) if the analysis concentration is below the trip point, if the analysis concentration is above the trip point, then an alarm condition will be created or maintained.

FAILSAFE: A non-defeated alarm that is in FAILSAFE mode energizes an alarm relay in a non-alarm condition and de-energizesan alarm relay in an alarm condition.

Note: Failsafe condition of an alarm is in software. This is not related to relays that have both normally-open and normally-closed terminals.

Teledyne Analytical Instruments

53

Image 53
Contents Series Teledyne Analytical Instruments SeriesTeledyne Analytical Instruments Iii Safety MessagesSeries Combustible GAS Usage Additional Safety InformationSeries Teledyne Analytical Instruments Vii Procedure for Removal of Internal Inaccessible Shock HazardsViii Table of Contents Operation Maintenance & Troubleshooting Appendix B AppendixTeledyne Analytical Instruments Xiii List of FiguresList of Tables Series Introduction Main Features of the AnalyzerIntroduction Introduction Series Principle of OperationAnalyzer Description Model 4060 Front Panel Interface Applications Introduction Operational TheoryModes of Operation Series Operational TheorySample System Analyzer SubsystemsOperational Theory Series Gas Flow Control System Internal Temperature ControllersFlame Ionization Detection Cell Fuel and Blanket Air SystemsTypical Piping Diagram for the Model Detection Cell Flame Ionization Cell Electrometer-AmplifierFlame Ignition Circuit Anode Power SupplyFlame Guard Circuit Series Installation InstallationUnpacking the Analyzer Mounting the AnalyzerElectrical Power Connections Installation SeriesUser Connections Electronic ConnectionsModel 4060 Rear Panel with Optional Gas Manifold Equipment Interface Connector Pin Arrangement Pin Function Threshold Alarm System Alarm Remote Calibration Connections Range ID Relay Connections 2.10 RS-232 Port Pin out of 50 pin D-Sub ConnectorGas Connections Commands via RS-232 InputGas Connections Less than 0.5 PPM Placing the System in Operation Series Operation OperationEquipment Operation Series Preliminary Power-Off Check ListActivating the Support Gases 1 Air Flame IgnitionCarrier Gas Span GasAnalyzer Operation Sample PumpVerification of the Flame Guard Circuit Ignition and/or Flame Guard Circuit FailureSeries Operation Style Conventions Default ParametersEnter Navigation and Data EntryMenu Structure EscapeMain Menus Span Model Screen System Self-Diagnostic TestALT-SPAN Alarms Function AL-1 Defeated AL-1 High AL-1 NON-FAILSAFE AL-1 NON-LATCHING Operation Series Range Function Analog Output AdjustmentChanging Stream Setting up an AUTO-CALLinearization Autocal Password Timing Group SetupPassword Protection Entering a Password Enter NEW Password Screen Logging Out Repeat Password Entry ScreenStandby Series Operation Operation Series Teledyne Analytical Instruments Series Maintenance Maintenance & TroubleshootingAnode Voltage Check Maintenance SeriesMeasuring Circuit Electrical Checks Electronic Stability Printed Circuit Board ReplacementCollector Cable Temperature Control Electronic CheckIgnition and/or Flame Guard Circuit Checks Sampling System Printed Circuit Board Descriptions Flame Guard and Anode Power Supply PCBProportional Temperature Controller PCB Electrometer-Amplifier PCB Maintenance Series Series Appendix Specifications and Initial SettingsAppendix Analytical Instruments Recommended Spare Parts ListAppendix Series PC Board Assemblies Drawing ListAppendix Series Teledyne Analytical Instruments Appendix B B1 Addendum and Testing Results