Series 4060 | Maintenance | |
|
|
|
2.If no reading is obtained, disconnect the
WARNING: DO NOT TOUCH CAPACITOR C1 OR C2 OR THEIR RELATED CIRCUIT FOILS. A SHOCK HAZARD MAY EXIST.
3.Carefully remove the circuit board without touching any connections which might lead to C1 or C2. After removal, discharge the two capacitors by placing a jumper wire across each.
4.The anode voltage may also disappear or be greatly diminished when condensation inside the sensor has occurred, shorting the igniter to the sensor body across the wet insulator. This usually occurs when the flame is turned on, if the sensor has not been preheated for at least 1 hour.
5.1.2Electronic Stability
If the checks outlined above indicate that conditions are normal, allow the analyzer to run electronically with the collector cable disconnected for several hours in the lowest range, and with the zero offset value adjusted so that the recorder is reading midscale. If all is normal electronically, a
5.1.3 Printed Circuit Board Replacement
If performance is not adequate, then the analyzer must be recalibrated as described in Section 4.6.5: Span before being placed back in service.
Teledyne Analytical Instruments | 67 |