Teledyne 4060 manual Alarms Function

Page 52

Operation

Series 4060

 

 

 

 

To initiate a self-test:

<UP/DOWN >From the MAIN MENU scroll to the SELF-TEST function.

<Enter> Activate the SELF-TEST function by pressing <Enter>. This brings up the SELF-TEST initialization screen.

________________________________________________

Begin Self-Test?

ENTER=Yes ESCAPE=NO

________________________________________________

<Enter or Escape> Start the diagnostic testing routine by pressing <Enter> or cancel out by pressing <Escape>.

If you pressed <Enter> the self-test routine will begin and after a few moments the results will appear onscreen. The module is functioning properly if it is followed by GOOD otherwise it is followed by BAD. In this case, please contact TAI Customer Service for an explanation of the problem.

To return the analyzer to the MAIN MENU, press <Enter> after the results screen.

If you pressed <Escape> you will be returned to the Analyze Mode.

4.6.9 The Alarms Function

The Model 4060 is equipped with two alarms and a system failure alarm relay. Each alarm relay has a set of form "C" contacts rated for 3 amperes resistive load at 250 VAC. See Figure in Chapter 3, Installation and/or the Interconnection Diagram included at the back of this manual for relay terminal connections. The alarm relay contacts are accessible to the user from 50-pin Equipment Connector. See Section 3.3.2.5 and Table 3-2.

The system failure alarm has a fixed configuration described in Chapter 3 Installation.

The concentration alarms can be configured from the ALARM function screen as follows:

________________________________________________

52

Teledyne Analytical Instruments

Image 52
Contents Series Series Teledyne Analytical InstrumentsSafety Messages Teledyne Analytical Instruments IiiSeries Additional Safety Information Combustible GAS UsageSeries Procedure for Removal of Internal Inaccessible Shock Hazards Teledyne Analytical Instruments ViiViii Table of Contents Operation Maintenance & Troubleshooting Appendix Appendix BList of Figures Teledyne Analytical Instruments XiiiList of Tables Introduction Main Features of the AnalyzerSeries Introduction Analyzer Description Principle of OperationIntroduction Series Model 4060 Front Panel Interface Applications Operational Theory IntroductionModes of Operation Series Operational TheoryOperational Theory Series Analyzer SubsystemsSample System Internal Temperature Controllers Gas Flow Control SystemFuel and Blanket Air Systems Flame Ionization Detection CellTypical Piping Diagram for the Model Detection Cell Electrometer-Amplifier Flame Ionization CellFlame Guard Circuit Anode Power SupplyFlame Ignition Circuit Installation Series InstallationUnpacking the Analyzer Mounting the AnalyzerInstallation Series Electrical Power ConnectionsUser Connections Electronic ConnectionsModel 4060 Rear Panel with Optional Gas Manifold Equipment Interface Connector Pin Arrangement Pin Function Threshold Alarm System Alarm Remote Calibration Connections Range ID Relay Connections Pin out of 50 pin D-Sub Connector 2.10 RS-232 PortCommands via RS-232 Input Gas ConnectionsGas Connections Less than 0.5 PPM Placing the System in Operation Equipment OperationSeries Operation Activating the Support Gases Preliminary Power-Off Check ListOperation Series Flame Ignition 1 AirCarrier Gas Span GasSample Pump Analyzer OperationVerification of the Flame Guard Circuit Ignition and/or Flame Guard Circuit FailureSeries Operation Default Parameters Style ConventionsNavigation and Data Entry EnterEscape Menu StructureMain Menus Span ALT-SPAN System Self-Diagnostic TestModel Screen Alarms Function AL-1 Defeated AL-1 High AL-1 NON-FAILSAFE AL-1 NON-LATCHING Operation Series Analog Output Adjustment Range FunctionLinearization Setting up an AUTO-CALChanging Stream Autocal Password Password Protection Group SetupTiming Entering a Password Enter NEW Password Screen Repeat Password Entry Screen Logging OutStandby Series Operation Operation Series Teledyne Analytical Instruments Maintenance & Troubleshooting Series MaintenanceMeasuring Circuit Electrical Checks Maintenance SeriesAnode Voltage Check Printed Circuit Board Replacement Electronic StabilityTemperature Control Electronic Check Collector CableIgnition and/or Flame Guard Circuit Checks Sampling System Flame Guard and Anode Power Supply PCB Printed Circuit Board DescriptionsProportional Temperature Controller PCB Electrometer-Amplifier PCB Maintenance Series Appendix Specifications and Initial SettingsSeries Appendix Appendix Series Recommended Spare Parts ListAnalytical Instruments Drawing List PC Board AssembliesAppendix Series Teledyne Analytical Instruments B1 Addendum and Testing Results Appendix B