Teledyne 4060 manual Placing the System in Operation

Page 40

Installation

Series 4060

 

 

 

 

 

 

Addenda included with this manual to determine if special conditions apply.

3.3.3.1EFFLUENT

All the gases introduced into the detection cell vent from one fitting at the rear of the analyzer. TAI recommends that the cell be permitted to vent directly to the atmosphere wherever possible.

If a vent line is required, the installation must include a drop-out pot to collect the water that is formed by the burning of the hydrogen fuel. The vent line must be constructed so that water and dirt cannot collect in it.

3.3.3.2SAMPLE BYPASS VENT

The sample bypassed by the back-pressure regulation system vents from a separate port at the rear of the analyzer. If a vent line is required, it must be installed so that water and dirt cannot accumulate in it.

3.3.3.3FUEL AND AIR CONNECTIONS

The fuel used to provide combustion should be hydrogen gas, zero quality with certified THC of less than 0.5 ppm and supplied at a pressure of 40 psig. The compressed air, ultra zero gas quality with THC less than 0.1 ppm should be used and supplied at a pressure of 40 psig. Connect the fuel and air sources to the instrument according to the gas connection diagram included at the back of this manual.

3.3.3.4SAMPLE AND SPAN GAS CONNECTIONS

The sample and span gas connections are made at the rear panel. If the optional autocal module is installed, the sample and span gas will connect to this module. If no autocal module is present, these gases connect to the labeled fittings on the rear panel.

Set the supply pressure for sample and span gas to 20 psig.

3.4 Placing the System in Operation

See Section 4 for information on starting the analyzer for the first time. Makes sure that all electrical connections have been made correctly and all connectors are fully seated. Make sure all gas connections are correct and leak–free.

40

Teledyne Analytical Instruments

Image 40
Contents Series Series Teledyne Analytical InstrumentsSafety Messages Teledyne Analytical Instruments IiiSeries Additional Safety Information Combustible GAS UsageSeries Procedure for Removal of Internal Inaccessible Shock Hazards Teledyne Analytical Instruments ViiViii Table of Contents Operation Maintenance & Troubleshooting Appendix Appendix BList of Figures Teledyne Analytical Instruments XiiiList of Tables Introduction Main Features of the AnalyzerSeries Introduction Analyzer Description Principle of OperationIntroduction Series Model 4060 Front Panel Interface Applications Operational Theory IntroductionModes of Operation Series Operational TheoryOperational Theory Series Analyzer SubsystemsSample System Internal Temperature Controllers Gas Flow Control SystemFuel and Blanket Air Systems Flame Ionization Detection CellTypical Piping Diagram for the Model Detection Cell Electrometer-Amplifier Flame Ionization CellFlame Guard Circuit Anode Power SupplyFlame Ignition Circuit Installation Series InstallationUnpacking the Analyzer Mounting the AnalyzerInstallation Series Electrical Power ConnectionsUser Connections Electronic ConnectionsModel 4060 Rear Panel with Optional Gas Manifold Equipment Interface Connector Pin Arrangement Pin Function Threshold Alarm System Alarm Remote Calibration Connections Range ID Relay Connections Pin out of 50 pin D-Sub Connector 2.10 RS-232 PortCommands via RS-232 Input Gas ConnectionsGas Connections Less than 0.5 PPM Placing the System in Operation Equipment OperationSeries Operation Activating the Support Gases Preliminary Power-Off Check ListOperation Series Flame Ignition 1 AirCarrier Gas Span GasSample Pump Analyzer OperationVerification of the Flame Guard Circuit Ignition and/or Flame Guard Circuit FailureSeries Operation Default Parameters Style ConventionsNavigation and Data Entry EnterEscape Menu StructureMain Menus Span ALT-SPAN System Self-Diagnostic TestModel Screen Alarms Function AL-1 Defeated AL-1 High AL-1 NON-FAILSAFE AL-1 NON-LATCHING Operation Series Analog Output Adjustment Range FunctionLinearization Setting up an AUTO-CALChanging Stream Autocal Password Password Protection Group SetupTiming Entering a Password Enter NEW Password Screen Repeat Password Entry Screen Logging OutStandby Series Operation Operation Series Teledyne Analytical Instruments Maintenance & Troubleshooting Series MaintenanceMeasuring Circuit Electrical Checks Maintenance SeriesAnode Voltage Check Printed Circuit Board Replacement Electronic StabilityTemperature Control Electronic Check Collector CableIgnition and/or Flame Guard Circuit Checks Sampling System Flame Guard and Anode Power Supply PCB Printed Circuit Board DescriptionsProportional Temperature Controller PCB Electrometer-Amplifier PCB Maintenance Series Appendix Specifications and Initial SettingsSeries Appendix Appendix Series Recommended Spare Parts ListAnalytical Instruments Drawing List PC Board AssembliesAppendix Series Teledyne Analytical Instruments B1 Addendum and Testing Results Appendix B