Series 4060 | Operation | |
|
|
|
Operation
This section of the manual describes how to setup and operate the Model 4060 Analyzer. Sections 4.1 through 4.5 describe preliminary steps and equipment needed to operate the analyzer. Beginning with Section 4.6, the actual operation of the analyzer is described along with descriptions of the display prompts, messages and options available to the user within a menu or sub menu. You should read this chapter in its entirety and become familiar with the operating characteristics of this system before starting the analyzer for the first time. The Addendum and Testing Results section of Appendix B lists the software revision and default settings for your specific system.
4.1 Equipment
The following supporting gases and hardware will be required to operate the (standard) analyzer:
1.Fuel: A cylinder containing a hydrogen gas, zero gas quality composition will be required to supply the fuel for the flame ionization burner. The cylinder is to be equipped with an
2.Blanket Air: A cylinder of compressed air, zero gas quality will be required to maintain the proper atmosphere within the cell. The cylinder is to be equipped with an
3.Carrier Gas: A cylinder of nitrogen gas, zero gas quality, equipped with dual stage metallic diaphragm regulator is required.
4.Span Gas: A cylinder or a permeation device system, capable of generating known concentration of benzene and/or other aromatic hydrocarbons of interest, will be required to standardize the analyzer.
5.Sample Pressure Regulation: An